U2722A/U2723A Service Guide
3
Performance Verification and Adjustment
Performance Verification and Adjustment
This section contains the performance verification and
adjustment procedures. If the U2722A/U2723A fails any of
the tests or if any abnormal test results are obtained, return
the unit to the Keysight Service Center for readjustment.
Readjustment can also be carried out using Keysight
Measurement Manager revision 1.7 or above.
Recommended Test Equipment
The recommended test equipment for the performance
verification is listed in the table below. If the exact
equipment is not available, substitute the performance
verification standards of equivalent requirement(s).
Test Consideration
For optimum performance, all procedures should comply
with the following recommendations.
•
Ensure that the ambient temperature is stable and
between 22
°
C and 28
°
C. The ideal temperature for
performing performance verification should be 25
°
C
±
1
°
C.
•
Make sure that the ambient relative humidity is less than
80%.
•
Allow a three- hour warm
-
up period before performance
verification.
•
Keep the test connection cables as short as possible.
Table 3-1
Recommended test equipment
Equipment
Requirement(s)
Recommended model
Purpose
Used
Digital multimeter
(DMM)
•
8½ digits resolution
Keysight 3458A
Measures DC
voltage and
current
Performance
verification and
adjustment
Содержание u2722a
Страница 1: ...Keysight U2722A U2723A USB Modular Source Measure Units Service Guide ...
Страница 2: ......
Страница 3: ......
Страница 12: ...X U2722A U2723A Service Guide Contents Source Current Mode Programming and Readback Accuracy 25 Adjustment 26 ...
Страница 22: ...8 U2722A U2723A Service Guide 1 Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 48: ...34 U2722A U2723A Service Guide Index THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...