Features and Functions
2
Keysight U1701B User’s and Service Guide
33
These initial sets can be modified, please refer to the
Setting HI/LO limits
for
detailed description. Besides that, the power-on option can be used to restore the
default factory settings.
The following procedures will guide you on how to set compare mode:
1
Press
key momentarily to enter HI/LO Audible Alert mode. The
meter locks range in this mode. The
annunciator will be indicated, and
the secondary display will indicate
C01
to
C25
or the previously saved set
during last operation. The first left digit means comparison mode. The last
two digits indicate current comparison set. The primary display will indicate
the present measurement. The meter is ready to test.
2
Press
key momentarily to select different sets. The secondary
display will indicate
C01
to
C25
according to which comparison set has
been selected. Press and hold
SAVE
key for more than 1 s to save
comparison set for next entry.
3
Press
HI/LO
key to cycle through the HI/LO value used for comparison
and back to ready mode. The HI/LO limits will be briefly indicated on the
primary display, and then back to the ready mode.
4
If the reading is out of the HI/LO limits, the beeper will beep three times
and the secondary display will indicate
nGo
. If the reading is within the
HI/LO limits, the beeper will beep once and the secondary display will
indicate
Go
. After 3 s, the meter will return to ready state.
5
Press and hold the
key for more than 1 s to exit audible alert
mode.
Содержание u1701b
Страница 1: ...Keysight U1701B Dual Display Handheld Capacitance Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1701B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...1 Getting Started 22 Keysight U1701B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 25: ...Features and Functions 2 Keysight U1701B User s and Service Guide 25 Figure 2 1 Tolerance operation ...
Страница 28: ...2 Features and Functions 28 Keysight U1701B User s and Service Guide Figure 2 3 Relative Zero operation ...
Страница 31: ...Features and Functions 2 Keysight U1701B User s and Service Guide 31 Figure 2 4 Tolerance operation ...
Страница 34: ...2 Features and Functions 34 Keysight U1701B User s and Service Guide Figure 2 5 Set compare mode ...
Страница 37: ...Features and Functions 2 Keysight U1701B User s and Service Guide 37 Figure 2 7 Setting High Low limits ...
Страница 39: ...Features and Functions 2 Keysight U1701B User s and Service Guide 39 Figure 2 8 Capacitance measurement ...