Adjustment Procedures

 

AC 40 A range

1. 

Lock the RANGE button of the U1583B current clamp to 40 A.

2. 

Connect the current clamp to a multimeter as shown in Figure 1.

3. 

Set the multimeter to measure ACV with range > 400 mV full scale. 

4. 

Open the jaws of the current clamp and align to the 50 turns coil.

5. 

Connect calibrator AUX terminals to 50 turns coil. Configure calibrator to generate a 

 

current of 0.4 A with a 55 Hz frequency (equivalent to 20 A, 55 Hz with 50 turns coil) for the 

adjustment of the current clamp.

6. 

Remove two Phillips screws on the back of the current clamp and proceed to adjust VR1 using 

a cermanic trimmer until the display on the multimeter indicates AC 200 mV ± 0.2 mV. Refer to 
Figure 4 for the position of VR1.

 

 

      

Figure 4: U1583B Current Clamp Circuit Board Diagram

 

AC 400 A range

1. 

Release the Range button of the U1583B current clamp to 400 A.

2. 

Connect the output BNC of the U1583B to a BNC-to-dual banana converter plug and proceed to  

connect it to the output of the V (HI) and COM (LO) terminals of the multimeter.

3. 

Set the multimeter to AC 500.0 mV or 1000.0 mV.

4. 

Open the jaws of the current clamp and align to the 50 turns coil.

5. 

Connect calibrator 20 A and AUX LO terminal to 50 turns coil. Configure calibrator to generate a 

 

current of 4 A with a 55 Hz frequency (equivalent to 200 A, 55 Hz with 50 turns coil) for the 

 

adjustment of the current clamp.

6. 

Remove two Phillips screws on the back of the current clamp and proceed to adjust VR2 until the 

display on the multimeter indicates AC 200 mV ± 0.2 mV. Refer to Figure 4 for the position of VR2.

Remember to replace the screws to their original position after performing the calibration 

adjustments.

Maintenance

Repair or service should only be performed by a qualified personnel.

To avoid electrical shock or damage to the current clamp, do not allow moisture to get 

inside the case and remove all connections before opening the case.

 

 

Cleaning

 

-

Periodically wipe the case with a damp cloth and mild detergent. Do not use cleaners or solvents.

 

-

Open the jaws and wipe the metal areas of the jaws with a lightly oiled cloth, and then wipe the oil 

with a dry cloth. Do not allow rust or corrosion to form on the metal surface of the jaws.

 

Troubleshooting

 

If the current clamp does not perform properly, follow the steps below to identify the problem:

1. 

Inspect the mating surface of the jaws for cleanliness. The presence of foreign material may 

 

prevent the jaws from closing properly and affects the measurement results.

2. 

Verify and ensure the current clamp range is appropriate.

3. 

Verify and ensure the function and range selection on the multimeter or oscilloscope are 

appropriate.  

       VR1 for AC 40 A 

       adjustment

       VR2 for AC 400 A 

       adjustment

© Keysight Technologies 2009-2020

Edition 7, September 2020

Printed in Malaysia

U1583-90107

www.keysight.com

Содержание U1583B

Страница 1: ...agement Agency of Australia This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992 This instrument complies with the WEEE Directive 2002 96 EC marking requirement This affixed product label indicates that you must not discard this electrical electronic product in domestic household waste Safety Information Please use the Keysig...

Страница 2: ...vided by the equipment may be impaired Do not operate the current clamp around flammable gases or fumes vapor or wet environments Do not connect to the BNC output or the banana plug to any power sources Use proper terminals function and range for your measurements Standard Items Purchase Checklist The following items are included when you make a purchase U1583B current clamp BNC to banana plug ada...

Страница 3: ...der to meet the accuracy specifications Make sure the current clamp is perpendicular to the conductor Current Range Selection Release the yellow button for 40 A range and lock the button for 400 A range Button Range Output Release 40 A 10 mV A Lock 400 A 1 mV A Operation AC current can be measured without removing the conductor out of the circuit by following the procedure shown below 1 Connect th...

Страница 4: ... 34450A or equivalent 50 Turns Current Coil 0 2 A to 20 5 A Not applicable Fluke 5500A COIL or equivalent A one year interval is adequate for most applications Accuracy specifications are war ranted only if adjustment is made at regular calibration intervals Accuracy specifications are not warranted beyond the one year calibration interval Keysight does not recommend extending calibration interval...

Страница 5: ...0 A 55 Hz with 50 turns coil for the adjustment of the current clamp 6 Remove two Phillips screws on the back of the current clamp and proceed to adjust VR2 until the display on the multimeter indicates AC 200 mV 0 2 mV Refer to Figure 4 for the position of VR2 Remember to replace the screws to their original position after performing the calibration adjustments Maintenance Repair or service shoul...

Страница 6: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Keysight U1583B ...

Отзывы: