background image

Electrical Specifications

Range

Output 

Sensitivity

Accuracy      

 

± (% of r as specified below) at 23 °C ± 5 °C, with relative humidity less than 

80% RH

Span

48 Hz to 65 Hz

40 Hz to 48 Hz/

65 Hz to 1 kHz

[1]

1 kHz to 10 kHz

[1]

40 A

10 mV/A

0.5 A ~ 40 A

2 % + 0.5 A

5 % + 0.5 A

10 % + 0.5 A

400 A

1 mV/A

5 A ~ 40 A

2.5 % + 0.5 A

4.5 % + 0.5 A

12.5 % + 0.5 A

40 A ~ 200 A

2 % + 0.5 A

4 % + 0.5 A

12 % + 0.5 A

200 A ~ 400 A

1.5 % + 0.5 A

3.5 % + 0.5 A

11.5 % + 0.5 A

[1] Typical accuracy applies across all frequency ranges above 440Hz.

• The current conductor must be centered within the jaw aperture and no influence from 

adjacent currents.

• Connect to load impedance > 1 MW ||100 pF. 

• Crest factor < 3.

Current Clamp with

BNC connector

BNC-to-banana 

 plug adapter

Multimeter

Current Clamp with

BNC connector

Oscilloscope

  

  Figure 1: Connection Diagram

Alignment Marks

Put only one conductor within the jaws and align the conductor to the 

indicated marks as much as possible in order to meet the accuracy  

specifications. Make sure the current clamp is perpendicular to 

 

the conductor.

Current Range Selection

Release the yellow button for 40 A range and lock the button for 400 A range.

Button

Range

Output

Release  

~ 40 A

~ 10 mV/A

Lock  

~ 400 A

~ 1 mV/A

Operation

AC current can be measured without removing the conductor out of the circuit by following the procedure 

shown below:
1. 

Connect the current clamp to a multimeter or oscilloscope as shown in Figure 1.

2. 

Set the multimeter to measure ACV with measurement range greater than 400 mV full scale.

3. 

Position and align the jaw to a single conductor accordingly.

4. 

Ensure that the arrow marked on the clamp jaw points towards the load for phase measurements, or   

away from the load (toward the source) for neutral measurements.

5. 

Observe the AC value on the multimeter or the waveform on the oscilloscope which is proportional to 

 

the current.

6. 

Select a lower range on the current clamp and set the corresponding sensitivity (mV/A setting) on the 

oscilloscope if required.

Figure 2: Allignment Marks

Figure 3: Conductor Allignment Direction

Содержание U1583B

Страница 1: ...agement Agency of Australia This signifies compliance with the Australia EMC Framework regulations under the terms of the Radio Communication Act of 1992 This instrument complies with the WEEE Directive 2002 96 EC marking requirement This affixed product label indicates that you must not discard this electrical electronic product in domestic household waste Safety Information Please use the Keysig...

Страница 2: ...vided by the equipment may be impaired Do not operate the current clamp around flammable gases or fumes vapor or wet environments Do not connect to the BNC output or the banana plug to any power sources Use proper terminals function and range for your measurements Standard Items Purchase Checklist The following items are included when you make a purchase U1583B current clamp BNC to banana plug ada...

Страница 3: ...der to meet the accuracy specifications Make sure the current clamp is perpendicular to the conductor Current Range Selection Release the yellow button for 40 A range and lock the button for 400 A range Button Range Output Release 40 A 10 mV A Lock 400 A 1 mV A Operation AC current can be measured without removing the conductor out of the circuit by following the procedure shown below 1 Connect th...

Страница 4: ... 34450A or equivalent 50 Turns Current Coil 0 2 A to 20 5 A Not applicable Fluke 5500A COIL or equivalent A one year interval is adequate for most applications Accuracy specifications are war ranted only if adjustment is made at regular calibration intervals Accuracy specifications are not warranted beyond the one year calibration interval Keysight does not recommend extending calibration interval...

Страница 5: ...0 A 55 Hz with 50 turns coil for the adjustment of the current clamp 6 Remove two Phillips screws on the back of the current clamp and proceed to adjust VR2 until the display on the multimeter indicates AC 200 mV 0 2 mV Refer to Figure 4 for the position of VR2 Remember to replace the screws to their original position after performing the calibration adjustments Maintenance Repair or service shoul...

Страница 6: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Keysight U1583B ...

Отзывы: