Performance Tests and Calibration
7
Keysight U1401B User’s and Service Guide
149
Adjustment Procedures
Input calibration
1
Set the slide switch to the
M/S
position.
2
Allow the instrument to warm up for 20 minutes before performing calibration.
3
To enter the calibration mode, press
and
for more than one
second. The primary display will indicate
CHEEP
.
4
Press
to enter the input calibration mode.
5
Turn the rotary switch to the “
Test Function
” position as shown in the
adjustment table.
6
The primary display will show the corresponding reference value of the Cal
item.
7
Configure each Cal item.
8
Use the
or
arrow keys to select the Cal range.
9
Apply the input signal shown in the Input column of the table.
10
The secondary display will now show the value of the input voltage that is
needed; press the HOLD button to switch between the Offset and Gain values.
The secondary display will now show the Offset or Gain values.
11
Press
to complete the calibration tests.
12
Repeat steps 5 through 10 for each adjustment point.
13
To exit the calibration mode, press
and
for more than one second.
Содержание U1401B
Страница 1: ...Keysight U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 10: ...10 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 61: ...Making Measurements 3 Keysight U1401B User s and Service Guide 61 Figure 3 1 DC voltage measurement ...
Страница 76: ...3 Making Measurements 76 Keysight U1401B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 107: ...Application Examples 5 Keysight U1401B User s and Service Guide 107 Figure 5 5 Zener diode test ...
Страница 128: ...5 Application Examples 128 Keysight U1401B User s and Service Guide Figure 5 17 Verifying a frequency transmitter ...