Calibration Procedures
1
Keysight U1281A/U1282A Service Guide
47
DC V
Short
SHort
Zero reference point
6.00 V
3.00 V
0.9 to 1.1 x Reference value
60.00 V
30.00 V
0.9 to 1.1 x Reference value
600.00 V
300.00 V
0.9 to 1.1 x Reference value
1000.00 V
300.00 V
0.9 to 1.1 x Reference value
Short
SHort
Short V
Ω
/ COM terminals
DC mV
Short
SHort
Zero reference point
60.00 mV
30.00 mV
0.9 to 1.1 x Reference value
600.00 mV
300.00 mV
0.9 to 1.1 x Reference value
Short
SHort
Short V
Ω
/ COM terminals
AC mV
60.00 mV
3.00 mV (55 Hz)
Reference value
30.00 mV (55 Hz)
0.9 to 1.1 x Reference value
60.00 mV (55 Hz)
0.9 to 1.1 x Reference value
30.00 mV (1 kHz)
0.9 to 1.1 x Reference value
30.00 mV (10 kHz)
0.9 to 1.1 x Reference value
600.00 mV
30.00 mV (55 Hz)
Reference value
300.00 mV (55 Hz)
0.9 to 1.1 x Reference value
600.00 mV (55 Hz)
0.9 to 1.1 x Reference value
300.00 mV (1 kHz)
0.9 to 1.1 x Reference value
300.00 mV (10 kHz)
0.9 to 1.1 x Reference value
Temperature
[b]
Type-K
0.0 ºC
0.0 ºC with ambient compensation
Table 1-19
Adjustment input values (continued)
Test function
Step
Reference value
Valid reference input
Содержание U1281A
Страница 1: ...Keysight U1281A U1282A Handheld Digital Multimeter Service Guide ...
Страница 14: ...14 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 58: ...1 Calibration Procedures 58 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 68: ...2 Service and Maintenance 68 Keysight U1281A U1282A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...