Square Wave Output (for U1252B)
The square wave output function can be used to generate a
PWM (pulse width modulation) output or provide a
synchronous clock source (baud rate generator). You can
also use this function to check and calibrate flow- meter
displays, counters, tachometers, oscilloscopes, frequency
converters, frequency transmitters, and other frequency
input devices.
1
Turn the rotary switch to
position. The default
display setting is 600 Hz on the secondary display and
50% duty cycle on the primary display.
2
Press
or
to scroll through the available frequencies
(choose from the available 28 frequencies):
3
Press
to select the duty cycle (%) on the primary
display.
4
Press
or
to adjust the duty cycle. Duty cycle can be
set for 256 steps and each step is 0.390625%. The display
only indicates the best resolution with 0.001%.
Frequency (Hz)
0.5, 1, 2, 5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600,
2400, 4800
N O T E
Pressing
is the same as pressing
.
Содержание U1251B
Страница 1: ...Keysight U1251B and U1252B Handheld Digital Multimeter User s and Service Guide ...
Страница 2: ......
Страница 3: ...Keysight U1251B U1252B User s and Service Guide I ...
Страница 12: ...X Keysight U1251B U1252B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...XVI Keysight U1251B U1252B User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 140: ...118 Keysight U1251B U1252B User s and Service Guide 5 Maintenance THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 196: ...174 Keysight U1251B U1252B User s and Service Guide 7 Specifications THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...