Performance Tests and Calibration
6
Keysight U1251B/U1252B User’s and Service Guide
125
Charge terminal alert test
This test determines if the charge terminal alert is operating
properly.
The meter sounds an alert when the
terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the
position. The meter sounds
an alert beep and the primary display flashes the “Ch.Err”
until the lead is removed from the
terminal.
Figure 6-3
Charge terminal alert
N O T E
Before conducting this test, make sure the beep function is not disabled at
setup.