Performance Tests and Calibration
6
Keysight U1211A, U1212A, and U1213A User’s and Service Guide
101
Valid adjustment reference input values
Adjustments can be performed using the following reference input values:
Table 6-5
U1211A valid adjustment reference input values
Function
Range
Reference input value
Valid range for reference input
DC voltage
Short
SHORT
Short
V
and
COM
terminals
400 V
300.0 V
0.9 to 1.1
×
reference input value
1000 V
1000 V
0.9 to 1.1
×
reference input value
AC voltage
400 V
030.0 V (70 Hz)
0.9 to 1.1
×
reference input value
300.0 V (70 Hz)
0.9 to 1.1
×
reference input value
300.0 V (2 kHz)
0.9 to 1.1
×
reference input value
1000 V
100 V (70 Hz)
0.9 to 1.1
×
reference input value
1000 V (70 Hz)
0.9 to 1.1
×
reference input value
1000 V (2 kHz)
0.9 to 1.1
×
reference input value
AC current
40 A
02.00 A (70 Hz)
0.9 to 1.1
×
reference input value
30.00 A (70 Hz)
0.9 to 1.1
×
reference input value
400 A
030.0 A (70 Hz)
0.9 to 1.1
×
reference input value
300.0 A (70 Hz)
0.9 to 1.1
×
reference input value
1000 A
50 A (70 Hz)
0.9 to 1.1
×
reference input value
300 A (70 Hz)
0.9 to 1.1
×
reference input value
Resistance
Short
SHORT
Short
Ω
and
COM
terminals
4 k
Ω
3.000 k
Ω
0.9 to 1.1
×
reference input value
400
Ω
300.0
Ω
0.9 to 1.1
×
reference input value
Capacitance
400 µF
300.0 µF
0.9 to 1.1
×
reference input value
4000 µF
3000 µF
0.9 to 1.1
×
reference input value
Diode
Short
SHORT
0
Ω
2.000 V
2.000 V
0.9 to 1.1
×
reference input value
Содержание U1211A
Страница 1: ...Keysight U1211A U1212A and U1213A Clamp Meters User s and Service Guide ...
Страница 10: ...10 Keysight U1211A U1212A and U1213A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight U1211A U1212A and U1213A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight U1211A U1212A and U1213A User s and Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...