1
Calibration Procedures
40
Keysight U1190A Series Service Guide
Exiting the adjustment mode
1
Remove all the shorting plugs and connectors from the instrument.
2
Power off and on again.
Temperature
12.209 mV
-
-
-
✔
0.9 to 1.1
×
Reference value
45.119 mV
-
-
-
✔
0.9 to 1.1
×
Reference value
0 °C
-
-
-
✔
0.9 to 1.1
×
Reference value
[c] Ensure the clamp meter is turned on and stabilized for at least 1 hour with the thermocouple connected between the clamp meter and the
calibrator output terminal.
Vsense
(Hi.SE)
9 V, 60 Hz
-
✔
✔
✔
N/A
Vsense
(Lo.SE)
27 V, 60 Hz
-
✔
✔
✔
N/A
Table 1-8
Close-case adjustment inputs (continued)
Test function
Reference value
U1191A
U1192A
U1193A
U1194A Valid reference input