background image

This information is subject to change 
without notice. 
© Keysight Technologies 

2014-

2015

Edition 1, 

October 2015

N9030-90066

www.keysight.com

Содержание PXA N9030A

Страница 1: ...rom RF or Microwave to Millimeter Wave Frequency Installation Note NOTICE This document contains references to Agilent Technologies Agilent s former Test and Measurement business has become Keysight Technologies For more information go to www keysight com ...

Страница 2: ...ed to the implied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number N9030 90066 Edition Edition 1 October 2015 Printed in USA Malaysia Published by Keysight Technologies Inc 1400 Fo...

Страница 3: ...ed PXA N9030A Serial Numbers All To Be Performed By X Keysight Service Center Personnel Qualified by Keysight Customer Estimated Installation Time Estimated Adjustment Time Estimated Verification Time 6 0 Hours 8 5 Hours Varies with installed options Option Ordered Description N9030AK F04 PXA Upgrade from 3 6 GHz to 44 GHz N9030AK F05 PXA Upgrade from 3 6 GHz to 50 GHz N9030AK F08 PXA Upgrade from...

Страница 4: ...y frequency range labels to the instrument and perform the necessary adjustments and performance verification tests IMPORTANT Although the maximum frequency range of the instrument will be upgraded with this kit the instrument s optional internal preamp frequency will not be updated The option is licensed for one instrument model number serial number combination The license file that is downloaded...

Страница 5: ... J9 for instruments without MPB N9010 20006 1 W3 Cable semi rigid Cal In to attenuator Reference assembly J701 to Attenuator A N9020 20072 1 W20 Cable semi rigid Front End assembly J4 To YTO N9020 20152 1 W8 Cable assembly A11J2 to A13J2 Semi Rigid N9020 20131 1 W17 Cable semi rigid LO Out IF In to W29 option EXM N9020 20167 1 W30 Cable coax Ext IF Out A13J13 to Front End Control J903 8121 2025 1 ...

Страница 6: ...1 W61 Cable semi rigid YTF Output to Switch 4 Port 3 N9020 20129 1 W62 Cable semi rigid Switch 4 Port 2 to A13A1 J9 N9020 20130 1 Connector Body for switch cable 1252 1873 1 A11 Low band Switch N9020 60051 1 Low Band Switch Bracket N9020 00013 1 A12 YTF Preselector 5087 7326 or N9030 60024 1 A12MP1 Gap Pad 5022 7179 1 A13 Front End assembly N9030 60010 1 A15 Front End Controller N9020 60172 19 Scr...

Страница 7: ...ed above For example if the Product listed on the Option Upgrade Entitlement Certificate is N9030AK F09 the Product Number noted above should be N9030A 3 Press System Show Hardware Note the Part Matl Rev OF Rev and Hw Id of the Front End Controller in the table below 4 Refer to the data in step 3 above If the HW ID of the Front End controller is 41 it will be necessary to perform the Replace A15 F...

Страница 8: ...o support the millimeter wave instruments The hard drive re imaging procedure can be found at http mktwww srs is keysight com field service signal pxa hd_reimage index html Select the latest Windows XP WXP version in the list of available images License Installation Procedure over USB 1 Locate the Option Upgrade Entitlement Certificate 5964 5178 from the kit 2 Redeem the Option Upgrade Entitlement...

Страница 9: ...Installation Note N9030 90066 9 Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency Figure 1 USB Storage Device Configuration Menu ...

Страница 10: ...on Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer C Program Files Agilent licensing Verify the License Installation 1 Cycle power on the signal analyzer and wait until the analyzer boots to the measurement application screen The instrument will fail RF Alignment and the noise flo...

Страница 11: ...with the attenuator still attached Unplug the ribbon cable from both attenuators The ribbon cable will be used for the new attenuators Discard the attenuator bracket 6 Remove the input connector assembly with the W2 cable still attached by removing the two screws that attach the connector bracket to the instrument frame Discard 7 Locate the 2 4 mm input connector in the kit and install it on the i...

Страница 12: ...acket to the chassis and remove the Low Band Switch bracket Remove the A12 YTF assembly 20 Disconnect all semi rigid cables from the YTF If Option MPB LNP was present you already removed these cables Discard the cables 21 Disconnect the wire harness from the YTF header This wire harness will be reused 22 Remove the YTF by removing the 4 mounting screws on the instrument chassis If the gap pad is s...

Страница 13: ... N9020 00013 in the kit 33 Refer to the PXA Service Guide Chapter 16 RF Area Options 543 544 and 550 See figure 16 47 that shows how to mount Low Band Switch to Bracket 34 Refer to figure 16 46 in the service guide Place the switch bracket into place in the chassis Loosely install the first 0515 0372 screw closest to the front of the instrument 35 Locate the loose end of the W11 wire harness that ...

Страница 14: ...tor 33326 60013 and locate 2 0515 0372 screws 42 Refer to Figure 3 With the attenuator positioned with the ribbon cable connector pointing up place magnetic shield item 5 onto attenuator then place Attenuator bracket over magnetic shield align the mounting holes and secure together with 2 0515 0372 screws Torque to 9 inch pounds 43 Install the attenuator assembly into the instrument as shown in Fi...

Страница 15: ...his Front End assembly instruments WITHOUT Option EXM External Mixing require the load 48 See Figure 4 If the instrument contains Option EXM locate the semi rigid cable N9020 20167 in the kit This is W17 Connect the end with the SMA male connector to A13A1J8 with the female connector pointing towards where the front panel would be Torque to 10 inch pounds Figure 4 Orientation of W17 49 If the inst...

Страница 16: ...iginally contained these options Installing Switches and Control Cables 50 Refer to Figure 5 Locate the two switches and the switch bracket in the kit Figure 5 Bracket Mount to Switches 51 Attach the switches 2 and 4 to the bracket with the 0515 1934 screws Note that the switch with the wire pigtail 4 mounts below the switch without the pigtail 2 Torque to 6 inch lbs ...

Страница 17: ... wire harness that has the yellow black green wires into the bottom switch harness Figure 6 Switch Wiring 53 Locate the ribbon cable N9020 60177 3 in the kit and install it onto the top switch Note the cable must have a fold in it as shown Ribbon cable 4 is the existing W7 cable removed from the Low Band Switch assembly earlier Notice how this cable will be routed between the switches The other en...

Страница 18: ...e switch bracket assembly into the frame Route W7 ribbon cable between the switches Notice the position of the other ribbon cable W55 and wire harness W12 Secure the switch assembly with two 0515 1035 screws 1 Torque to 9 inch lbs Insert W7 ribbon cable into Low Band Switch header Ensure the locking tabs are engaged Figure 7 Cable Routing ...

Страница 19: ...de from RF or Microwave to Millimeter Wave Frequency Installing Semi Rigid Cables 55 Refer to Figure 8 Locate W52 N9020 20125 1 and W54 N9020 20127 2 in the kit and install them as shown Torque cables to 10 inch lbs Figure 8 W52 and W54 Cable Installation ...

Страница 20: ... Frequency 56 Refer to Figure 9 Locate W60 N9020 20128 1 in the kit and install it as shown Locate W61 N9020 20129 2 and install it as shown Locate W59 N9020 20151 3 and install it Locate W62 N9020 20130 4 and install it Torque cables to 10 inch lbs Figure 9 W60 W61 W62 and W59 Cable Installation ...

Страница 21: ...0131 2 in the kit and install between the A11 Low Band Switch J2 and the Front End assembly A13A1J2 Refer also to Figure 11 Torque to 10 inch lbs 59 Locate W20 N9020 20152 3 in the kit and install between the YTO and the Front End assembly A13A1J4 Torque to 10 inch lbs 60 Locate W3 N9020 20072 4 in the kit and install from Attenuator A Cal In and A16 Reference assembly J701 Torque to 10 inch lbs 6...

Страница 22: ...22 Installation Note N9030 90066 Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency Figure 11 A13 Front End Assembly Millimeter Wave ...

Страница 23: ...Press down on the W7 ribbon cable header on the Front End Control assembly to assure it did not become loose during the installation Also assure the ribbon cable is routed where it will not get damaged when the covers are replaced 64 Attach the wire harness W12 as shown in Figure 12 Assure the wire harness is routed so it will not get damaged when the covers are replaced 65 Attach ribbon cable W55...

Страница 24: ...o 10 inch pounds 68 Locate W2 N9020 20173 in the kit and install between Attenuator A and the input connector Torque to 10 inch lbs Complete the Hardware Installation 69 Replace the right side chassis and front frame assembly If Option EXM External Mixing is present reconnect the semi rigid cable behind the front panel EXT MIXER connector 70 Replace the power cord and power on the instrument to as...

Страница 25: ...tem and look for the Instrument S W Revision The latest revision of the X Series Signal Analyzers software may be downloaded from http www keysight com find xseries_software Contact the division for removal of old frequency range license 1 Connect the instrument to the Keysight LAN The division will need to log into the instrument and remove the License for the old frequency range 2 E mail the div...

Страница 26: ...abels are printed with the option number and frequency range Example 544 44 GHz 2 Remove the bar code portion of the label and discard this bar code portion Apply this label to the rear panel as shown in Figure 14 Figure 14 Rear Panel Label 3 Locate the correct front panel label that indicates the new frequency range of the analyzer The front panel labels show only the new frequency range Example ...

Страница 27: ...Installation Note N9030 90066 27 Frequency Range Upgrade from RF or Microwave to Millimeter Wave Frequency Figure 15 Front Panel Label ...

Страница 28: ...ts Required Perform all possible adjustments Performance Testing Required Perform all Performance Verification Tests For assistance contact your nearest Keysight Technologies Sales and Service Office To find your local Keysight office access the following URL or if in the United States call the following telephone number http www keysight com find assist 1 800 829 4444 8 am 8 pm ET Monday Friday ...

Страница 29: ...This information is subject to change without notice Keysight Technologies 2014 2015 Edition 1 October 2015 N9030 90066 www keysight com ...