Keysight PD1000A Power Device Measurement System Startup Guide
On State and Off State S-Parameter Characterization
Running the S-Parameter Characterization
--
Drain Voltage (V):
Enter the voltage for SMU Channel 2 (rear panel) of the
B2902A SMU. Typically, this connects to the drain of the FET or
collector of the IGBT.
--
Drain Current Limit (A):
Enter the Current Limit for SMU Channel 2 (rear
panel) of the B2902A SMU.
2
(Optional) To save set-up time, you can save all configuration settings and
general settings to a file that you can re-use to skip the configuration process
the next time you test a DUT that uses the same settings.
a
Click on
Settings
on the main screen.
b
Select
Save Settings to File
. This displays a file save dialog.
c
Use the dialog to browse to the directory where you want to store the
settings.
d
Type a file name for the settings file using one of the following formats:
--
IVCV.kts
- for IV-CV tests
--
ONSTATE.kts
- for on-state S-parameter tests
--
OFFSTATE.kts
- for off-state S-parameter tests
e
Click
Save
. This saves your current settings into a Keysight Test Settings file
(.kts).
Calibration
Calibrate the ENA with the appropriate PD1000A Test Fixture connected. Use the
Calibration Standards supplied with the test fixture. Refer to the
PD1000A Test
Fixture User Guide
or the PD1000A Help file for calibration details.
Run the Test/Measurement
1
Click the
Start
button (
) to start a measurement.
–
You can stop an active measurement by clicking the Stop (
) button.
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......