Keysight PD1000A Power Device Measurement System Startup Guide
IV and CV Device Characterization
Running the IV and CV Tests
B150xA Settings:
Specify the B150xA VISA address, timeout, and in which slots the B150xA
Measurement Resources (SMUs) are installed. For example, in the previous
graphic, the Control - Voltage Control SMU (an MCSMU) is installed as SMU4
(therefore it is entered as: SMU4:MC).
1
Instrument type:
Use this drop-down menu to select either
B1505A
or
B1506A
.
2
Instrument options:
Use this drop-down menu to select the SMU module configuration
option:
H21
,
H51
, or
H71
.
3
B150xA VISA Address:
VISA Address of the B150xA. See
“Install IO Libraries Suite on
4
B150xA TimeOut (s):
Maximum time that the PD1000A waits for the B150xA to respond
to commands. (Note that some measurements take two to three minutes; therefore set
the Timeout time accordingly (300 s min.)).
5
Workspace Name:
Enter the WorkSpace Name you selected when first starting the
EasyEXPERT software on the B150xA. (If necessary, re-run the EasyEXPERT software on
the B150xA to set up this value.
6
Do not close Workspace after Measurement:
Default is checked. The system does not
close the B150xA EasyEXPERT Workspace after completing tests.
7
Temperature Sensor available in this fixture:
Select this checkbox if your test fixture has
a temperature sensor.
8
Transistor Thermistor Channel:
Use this drop-down menu to select either thermistor
Channel 1
or
Channel 2
to monitor temperature.
9
Input HCSMU:
Enter the name of the HCSMU used for input, for example,
SMU3:HC
.
–
Available only with the H21 option.
–
Leave this field blank for options H51 and H71.
–
Look up the module name from the HCSMU input field in the
Module
Selector
tab of the Configuration dialog in EasyEXPERT. (
See “Configure
10
Input – UHCU:
Enter the name of the UHCU (Ultra High Current Unit) SMU, for
example,
SMU4:UHC1500
.
–
Use with B150xA options H51 and H71 (500 A or 1500 A) only.
–
Leave this field blank for option H21 (20 A).
–
Look up the module name from the
UHC Expander / Fixture
tab of the
Configuration dialog in EasyEXPERT. (
–
For the B1506A:
-- The UHCU uses the test chamber and two MCSMUs installed in the
mainframe.
-- For the H51 option, use the UHCU named:
SMU:n:UHC500
.
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......