40
Keysight PD1000A Power Device Measurement System Startup Guide
IV and CV Device Characterization
IV and CV Test Requirements
Import Utility Files
This section uses the setup files that you downloaded to a memory stick to
perform a one-time installation of various utility files. (See “Download Setup and
Configuration Files” on page 19.)
1
Insert the memory stick into the front panel.
2
In the tabs on the left of the EasyEXPERT screen, select
Application Test.
3
Click on the
Category...
drop-down menu.
4
Select the
Utility
checkbox.
5
Click on the
Library
drop-down menu.
6
Select
Import...
This opens a file selection dialog.
7
Browse to the location where you saved the PD1000A setup files.
8
Select the following PD1000A setup files:
– Gate Diode PD1000A.xtd
– Ic-Vge PD1000A.xtd
– Id-Vgs PD1000A.xtd
– Rg, Cgg for IGBT PD1000A.XTD
– Rg, Cgg for PowerMOSFET PD1000A.XTD
– Temperature Sensor 1.xtd
– Temperature Sensor 2.xtd
– UHC Selector Input Selection.xtd
9
Click
Open
. This imports the selected files and makes them available for tests,
thermal platform support, and ultra-high current (UHC) support.
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......