TDR (Time Domain Reflectometry) - Option 215
TDR Measurement Settings
An un-terminated cable (with NO perfect load at the end) will show faults that
appear to be beyond the end of the cable. These are NOT alias faults. These
faults appear as the signal reflects off the open at the end of the cable and
travels back down the cable toward the connection at the FieldFox.
Re-reflections are measured at the FieldFox as mirror images of the original
faults. The largest fault is the open end of the cable. To avoid confusion, set
the Stop distance shortly after that fault.
How to Prevent Undersampling
Undersampling is when the TDR measurement is inaccurate because the
frequency step size (resolution) is too large. The predominant symptom of this
is when the measured impedance is incorrect. For example, when measuring a
75
Ω
cable, a marker that is on a part of the trace without reflections shows
56
Ω
.
To prevent undersampling, set
TDR Sweep
to
Cable
. If undersampling still
occurs, set
TDR Sweep
to
Auto
. Then set
Auto Max Frequency
to a lower
value. Manually increase the Resolution so that the Start Frequency setting is
less than about 500 kHz.
Содержание N9913A
Страница 6: ...6 ...
Страница 56: ...56 Keysight N9927 90001 User s Guide CAT Cable and Antenna Test Mode 2 Port Insertion Loss Measurements ...
Страница 82: ...82 Keysight N9927 90001 User s Guide TDR Time Domain Reflectometry Option 215 TDR Measurement Settings ...
Страница 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 142: ...142 Keysight N9927 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Страница 262: ...262 Keysight N9927 90001 User s Guide IQ Analyzer IQA Option 351 CPU2 Only Other Features Available ...
Страница 286: ...286 Keysight N9927 90001 User s Guide Interference Analyzer SA Mode Option 236 Record Playback SA Option ...
Страница 292: ...292 Keysight N9927 90001 User s Guide Reflection Mode SA Models Option 320 Reflection Mode Settings ...
Страница 316: ...316 Keysight N9927 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 332: ...332 Keysight N9927 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 346: ...346 Keysight N9927 90001 User s Guide VVM Vector Voltmeter Mode Option 308 A B and B A Measurements ...
Страница 370: ...370 Keysight N9927 90001 User s Guide AM FM Metrics SA Only and Option 355 Interpreting AM FM Metrics Error Messages ...
Страница 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Страница 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Страница 528: ...528 Keysight N9927 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 544: ...544 Keysight N9927 90001 User s Guide Safety Considerations Certification and Compliance Statements ...
Страница 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Страница 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Страница 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Страница 555: ...555 Keysight Handheld Analyzers N99xxA User s Guide E Keysight Software End User Licensing Agreement EULA ...
Страница 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Страница 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 565: ......