450 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Fan Assembly
Fan Assembly
Removal
1.
Remove the instrument outer case. Refer to the
removal procedure.
2.
Remove the front frame. Refer to the
removal
procedure.
3. For Serial Number Prefix < 5605 Options 508, 513, 526
:
—
Refer to
. Remove the nine screws
(1)
(0515-0372) and
the two screws
(2)
(0515-1946) that attach the fan bracket to the
chassis.
Figure 15-98
Fan Bracket Removal
Serial Number Prefix < 5605 Options 508, 513, 526
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...