192 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
14.
The analyzer should read 322.5 MHz at
−
23 dBm
±
3 dB as shown in
Figure 6-2
A15 322.5 MHz Output
Reconnect W25 to A15J900.
IMPORTANT
Measuring at this location is for convenience. The 322.5 MHz signal is an
output at A15J900. Since the W25 cable connects from this output to the
input of the Analog I.F. assembly at A2J100, the small coaxial cable has not
been tested at the A2J100 end.
If the 322.5 MHz signal is not measuring the correct power level, refer to
Section Troubleshooting (RF/Microwave Analyzers).”
in this service guide.
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...