156 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Low Band Preamp (Option P08, P13, P26)
The Preamp is aligned as part of the System Gain internal alignment process.
See the description of the initial alignments and the location of the alignment
history file in
Chapter 2, “Boot Up and Initialization Troubleshooting”, on page
. Viewing the Alignment History file will tell you if other alignment tests
failed, and reveal pass or fail information for the preamp.
If any of the preamp options are installed, the Low Band preamp can be verified
as follows:
Set up the analyzer to view the 50 MHz calibrator signal on screen
Press
AMPTD
,
Attenuation
, and verify that the Elec Atten is set to Disabled. It
is not possible to turn on both the electronic attenuator and preamp at the
same time.
To activate the internal preamplifier press
AMPTD
,
Signal Path
,
On
.
The expected operation is the signal level will not change more than
±
0.5 dB
and the noise floor will increase due to the fact the input attenuator is
automatically increased. If the amplitude error when switching the preamp on
and off is greater than 0.5 dB, assure that any other internal alignment failures
are resolved since the System Gain alignment must run before optimum
performance is possible. Assure you have evaluated all causes including A15
Front End Controller switching problems, before you replace the Front End
assembly.
If you determine the problem is at a frequency other than 50 MHz, perform the
low band flatness adjustment before replacing the Front End assembly.
Low Band Mixing Equations
Low Band first mixer:
RF = LO
−
IF
IF = LO
−
RF
LO = IF + RF
where RF is the input signal at the signal analyzer
Low Band second mixer:
RF = LO + IF
IF = RF
−
LO
LO = RF
−
IF
where RF in this case is the first IF signal (signal at A13A1J11)
Содержание N9040B Option 513
Страница 4: ...4 ...
Страница 138: ...138 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Messages Condition Messages ...
Страница 240: ...240 Chapter 11 Hardware Options A21 Wideband Analog IF Troubleshooting Figure 11 1 Analog IF Block Diagram ...
Страница 256: ...256 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 306: ...306 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 2 External Hardware ...
Страница 338: ...338 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts Figure 14 28 Chassis ...
Страница 346: ...346 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Replaceable Parts ...
Страница 502: ...502 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Post Repair Procedures ...
Страница 532: ...532 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Functional Tests Scale Fidelity ...
Страница 536: ...536 Keysight Technologies N9040B UXA Signal Analyzer Service Guide Instrument Software Software Updates ...
Страница 541: ...Index Index 541 Y yellow LED 48 51 yig filter see A12 YTF YTF see A12 YTF ...