33.
. Remove W20 from J710 of the A16 Reference assembly
(1)
.
34.
Using the number “6” provided (7121-5606), change the number on the end of W20 that was
removed from J710 of the A16 Reference assembly
(1)
from “710” to “716”.
35.
Reconnect W20 to J716 of the A16 Reference assembly
(1)
instead of J710.
36.
Moving the connection point of W20 on the A16 Reference assembly will leave extra slack in
the cable. Pull this slack down to the side of the instrument.
Figure 11