Installation Note N9020-90272

5

Option YAS Installation Kit

Contents

Tools Required for Installation

— Personal computer with internet access and USB port
— USB storage device with >2GB free memory

Tools Required for Manual Operation Verification

— Voltmeter
— Test cable, BNC (used to connect voltmeter to the analyzer rear panel ANALOG OUT)
— Banana plug to BNC adapter (used to connect voltmeter to BNC cable)

Installation Procedure over USB

1.

Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit.

2.

Redeem the Option Upgrade Entitlement Certificate by following the instructions on the 

Certificate.

3.

After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an 

attached License File.

4.

Locate a USB storage device. Perform a virus scan on this device before use.

5.

Save the License File to the root directory of the USB Storage Device.

6.

Connect the USB Storage Device to the signal analyzer USB port. Windows will detect the new 

hardware and may display the configuration menu shown in 

Figure 1

. This menu may be 

configured according to your preferences.

Quantity

Description

Keysight Part Number

1

Installation Note

This note

1

Option Upgrade Entitlement Certificate

5964-5178

Содержание N9010B EXA

Страница 1: ...Option YAS Y Axis Screen Video Out Keysight N9010B EXA and N9020B MXA Signal Analyzers N9048B PXE EMI Receiver Installation Note ...

Страница 2: ...lied warranties of merchantability and fitness for a particular purpose Keysight Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Manual Part Number N9020 90272 Edition Edition 1 October 2018 Supersedes February 2018 Printed in USA Malaysia Published by Keysight Technologies...

Страница 3: ...s Affected N9010B EXA Signal Analyzer N9020B MXA Signal Analyzer N9048B PXE EMI Receiver Serial Numbers All Options To Be Performed By X Keysight Service Center X Personnel Qualified by Keysight X Customer Estimated Installation Time Estimated Adjustment and Verification Time 0 5 Hours 0 5 Hours ...

Страница 4: ...ranted specifications Within one year of the initial calibration date of the instrument this option is fully calibrated with no further adjustment or verification testing To determine the initial calibration date locate the original calibration certificate that was shipped with the instrument at the time of purchase The Date of Calibration is printed on the original calibration certificate To ensu...

Страница 5: ... Redeem the Option Upgrade Entitlement Certificate by following the instructions on the Certificate 3 After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an attached License File 4 Locate a USB storage device Perform a virus scan on this device before use 5 Save the License File to the root directory of the USB Storage Device 6 Connect the USB Storage Device ...

Страница 6: ...nse Installation message similar to the one shown in Figure 2 Figure 2 Successful License Installation Alternate Installation Procedure The License File can be manually installed over USB or LAN by placing the license file in the following folder on the signal analyzer C Program Files Agilent licensing Verify the Installation 1 Cycle the power on the signal analyzer 2 Press System Show System to d...

Страница 7: ...dBm 3 Press AMPTD Ref Level and key in 25 dBm The displayed signal level should be at the top graticule line 4 Press Input Output Output Analog Out Screen Video 5 Connect a voltmeter set to measure DC volts to the analyzer rear panel ANALOG OUT connector 6 The voltmeter should measure 1V 7 Increase the Ref Level to 15 dBm The signal should drop one graticule division on screen and the voltmeter re...

Страница 8: ...red None A full calibration is required to assure the instrument meets all specifications The end user must ultimately determine whether they want a full calibration to be performed after the installation of this upgrade or not If a full calibration is required arrangements regarding the level of calibration must be made between the end user and the calibration provider For assistance contact your...