Functional Tests
Amplitude Accuracy at 50 MHz
Amplitude Accuracy at 50 MHz
Test Limits
Amplitude Accuracy should remain within 1.13 dB of the measured source
value across the range of source levels and changes in resolution bandwidth.
The Preamp (option P03, P07, P13, P26, P32, P44) should remain within ±1.3
dB of measured values.
A synthesized sweeper is used as the signal source for the test. The source
amplitude is varied using the signal source amplitude control. The attenuation
and resolution bandwidth are varied on the signal analyzer. The source
amplitude is measured by the power meter and signal analyzer at each setting,
and the values compared. The difference between each pair of measurements
indicates the amplitude accuracy.
Item
Critical Specifications
Recommended
Keysight Model
Adapter
a
Type-N (m), to 3.5 mm (m)
1250-1743
Adapter
3.5 mm (f) to 3.5 mm (f)
83059B
Adapter
b
2.4 mm (f) to 3.5 mm (m)
11901A
Adapter
2.4 mm (f) to 3.5 mm (f)
11901B
Attenuator, 30 dB
Accuracy: < 0.5 dB at 50 MHz
11708A
Cable
3.5 mm, 1 meter
11500E
Cable
BNC, 120 cm
10503A
Power Meter
Compatible with power sensor
E4418B
Power Sensor
Amplitude Range:
−
25 dBm to 10 dBm
8485A
Power Splitter
3.5 mm (f) connectors
6 dB loss
11667B
Synthesized Sweeper
Typical Temperature Stability: 0.01 dBc/
°
C
PSG
a. For N8973B, 74B, 75B
b. For N8976B
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...