310
Keysight NFA Series Noise Figure Analyzers Service Guide
Power Supply/Midplane Troubleshooting
A6 Power Supply Basic Troubleshooting (Cover On)
A6 Power Supply Basic Troubleshooting
(Cover On)
1.
Ensure the instrument is plugged into a known good AC power source and
the yellow standby LED near the power on switch is illuminated. A lit
yellow LED indicates the +5.1 SB VDC power supply voltage is providing
enough voltage to light the LED.
2.
Power on the instrument and verify that the green LED on the front panel
is lit. A lit LED indicates the power supply has received an “ON” command
from the A4, CPU assembly and that the +5.2A VDC supply is on.
3.
Verify both fans are operating on the side panel of the instrument. The fans
can be heard once the analyzer is powered up. If the fans are operating,
the fan power supply is functioning.
4.
The front panel probe power connector can be used to check the +15 VDC
(+/- 7%) and -12.6 VDC (+/-10%) supplies. The -12.6 VDC is produced by
post regulating the -15 VDC supply. A voltmeter can be used to verify
these supplies. Refer to
for a diagram of the probe power
connector.
Figure 10-2
Probe Power Connector
If these initial checks are functioning properly and yet the signal analyzer will
not boot up properly, or the display is not turning on, refer to
Up and Initialization Troubleshooting”
in this service manual.
There are no user replaceable fuses on the power supply. If the internal
fuse is blown, the power supply has experienced a major failure and should
be replaced. If you determine that the power supply is the failed assembly,
replace the power supply. If the power supply is suspect, perform the
following quick checks before removing the instrument cover.
Exercise great care when measuring voltages on the Probe Power
connector. Accidentally shorting the +15V or -12.6V supply to ground can
cause damage to the A1A2 or A1A5 assemblies.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...