uses 3rd-order for the N7744A and N7745A and 1st-order for the
N7747A and N7748A. The Drift D is calculated by
D = max
i
{F
i
} or mini{F
i
}
, whichever is worse.
Figure 43
Drift (dark)
Conditions:
All ports dark. Observation time as specified. Other
conditions as specified.
Dynamic range (logging mode)
For a fixed power meter range, dynamic range D specifies the difference
between the maximum displayable value P
max
in dBm in the
corresponding power meter range and the minimum measurable power
P
min
in dBm.
Conditions:
As specified.
Measurement:
P
max
is 3 dB above the nominal power meter range value
(that is, in the +0 dBm power meter range P
max
is +3 dBm), except for
the +10 dBm range with P
max
= +10 dBm. P
min
,dBm is calculated as the
root-sum-square (RSS) of the specified “Noise 2 sigma (dark)” (see,
“Noise 2 sigma (dark)” on page 66
and “Drift (dark)” (see,
. RSS is calculated in linear units (Watts).
Frequency Response
Specifies the frequency f
resp
up to which the power meter's transfer
function is above a defined cutoff level.
The transfer function is the measured optical amplitude on a modulated
sinusoidal optical input divided by the actual amplitude over
modulation frequency, expressed in dB.
Conditions:
Cutoff level as specified. Other conditions as specified.
4
Specifications and Regulations Compliance
64
Keysight N774x Series