background image

 

 

This information is subject to change 
without notice 
©Keysight Technologies, 2018 
December, 2018 
5992-3110EN 

www.keysight.com

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание N6731B

Страница 1: ...idesthedocumentation forthefollowinginstruments N6731B through N6736B N6741B through N6746B N6751A through N6756A N6761A through N6766A N6773A through N6777A N6781A through N6786A N6791A and N6792A Declassification and Security Document ...

Страница 2: ...cribed in this document are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcontract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1995 or as a commercial item as defined in FAR 2 ...

Страница 3: ...y Buyer shall pay all shipping charges duties and taxes Where to Find the Latest Information Documentation is updated periodically For the latest information about these products including instrument software upgrades application information and product information see the following URL https www keysight com en pc 851482 n6700 modular power system nid 35714 0 00 cc US lc eng To receive the latest...

Страница 4: ... Sales and Service Offices 5 2 Products Covered by this Document 6 3 Security Terms and Definitions 7 4 Instrument Memory 8 5 Summary of Memory Declassification Procedures 11 6 Procedure for Declassifying a Faulty Instrument 15 7 References 16 ...

Страница 5: ...the Internet at http www keysight com find assist If you do not have access to the Internet please contact your field engineer In any correspondence or telephone conversation refer to the instrument by its model number and full serial number With this information the Keysight representative can determine whether your unit is still within its warranty period ...

Страница 6: ...y the user in the instrument that needs to be saved is properly backed up before attempting to clear any of the instrument memory The model s user guide explains how to store states outside the product Keysight Technologies cannot be held responsible for any lost files or data resulting from the clearing or sanitization of memory Be sure to read this document entirely before proceeding with any fi...

Страница 7: ...or both Keysight declassification procedures are designed to meet the requirements specified in DoD 5220 22 M Chapter 8 Sanitization As defined in Section 8 301b of DoD 5220 22 M sanitization is the process of removing or eradicating stored data so that the data cannot be recovered using any known technology Instrument sanitization is typically required when an instrument is moved from a secure to...

Страница 8: ...and Size Writable During Normal Operation Data Retained When Powered Off Purpose Contents Data Input Method Location in Instrument and Remarks Clearing or Sanitization Procedure EEPROM Note 1 16K bit No Yes Used to store Model Number Serial Number Calibration data Calibration Date Programmed before installed and by Firmware operations cal routines Main PCA See Table 4 and See Table 5 FPGA 146KB 19...

Страница 9: ...ion Date Programmed before installed and by Firmware operations cal routines Main PCA See Table 4 and See Table 5 FPGA 1 4MB No No Power module measurement and control infrastructure Programmed During power up in Mainframe DPM Power Cycle 32M bit SPI Flash No No Memory not used for module when installed in mainframe Memory not used in module when installed in mainframe DPM Power Cycle Notes 1 When...

Страница 10: ...on in Instrument and Remarks Sanitization Procedure EEPROM 16K bit No Yes Used to store Model Number Serial Number Calibration data Calibration Date Programmed before installed and by Firmware operations cal routines Main PCA See Table 4 and See Table 5 FPGA 3 8MB No No Power module measurement and control infrastructure Programmed During power up in Mainframe DPM Power Cycle ...

Страница 11: ...of user data For Sanitizing N6700C N6701C and N6702C mainframes From the Front Panel Menu Select System Admin Login enter password Sanitize Enter or select Display will show Warning that all user data will be Sanitized and unit will be rebooted Press Enter or Select wait 15 seconds for unit to reboot before shutting power off The mainframe is now Sanitized When using remote interface send SYSTem S...

Страница 12: ...election 6 Erase any data displayed under date by pressing the back arrow key and press the Enter key 7 Press the Back key Select Save and the press the Enter key Press the Enter key again The unit will display Cal changes have been saved 8 Press the Back key three times 9 Select the Logout menu item and press the Enter key Press the Enter key again to log out 10 Repeat for all modules in the main...

Страница 13: ...or Login Logout enter 6 Select nonvolatile RAM Reset enter 7 Select Reset enter Note pressing enter clears all Non Volatile memory to factory defaults and the internal drive will be cleared 8 Wait for unit to prompt you to cycle power 9 After you cycle power the saved states for all modules in the mainframe will be reset to default values NV Ram is now cleared B To clear module calibration date fi...

Страница 14: ...ly module by pressing channel key 15 Enter the Administrator menu by pressing 16 Menu key 17 Select System enter 18 Select Admin enter 19 Select Login enter 20 Enter password enter 21 Enter 22 Select Cal enter 23 Select Date enter 24 Use the Back Space key to delete all the characters in the Date field and then press the Enter key 25 Press Back Key 26 Select Save enter 27 Select Save enter 28 Note...

Страница 15: ...formation cannot be cleared using the front panel or the remote interface The only choice in this situation is to take manual steps to remove any user information that may be present To manually remove any user information from the power module you must remove the module from the mainframe and destroy it ...

Страница 16: ...e downloaded in Acrobat PDF format from http www dss mil isp fac_clear download_nispom html 2 ISFO Process Manual for the Certification and Accreditation of Classified Systems under the NISPOM Defense Security Service DSS cleared industries may request a copy of this document via email by following the instructions at http www dss mil isp odaa request html ...

Страница 17: ...This information is subject to change without notice Keysight Technologies 2018 December 2018 5992 3110EN www keysight com ...

Отзывы: