14
When using the module in N6700B,N6701A,N6702A mainframes:
A.
To clear mainframe nonvolatile RAM, and module saved states,
1)
Turn on the mainframe.
2)
Enter the Admin menu by Pressing:
3)
The Menu key
4)
Select, System, enter
5)
Select, Admin, enter
6)
Login, enter
7)
Enter password , enter
8)
Enter,
9)
Select, Nvram, enter
10)
Select, Reset, enter
11)
Wait for the unit to prompt you to cycle power
12)
Cycle Power
Note: Upon power cycle the saved states for all modules in the mainframe will
be reset to default values
NV Ram is now cleared.
B.
To clear module calibration date field for each module.
13)
Turn on the mainframe.
14)
Select the appropriate power supply module by pressing channel key.
15)
Enter the Administrator menu by pressing
16)
Menu key
17)
Select, System, enter
18)
Select, Admin, enter
19)
Select, Login, enter
20)
Enter password , enter
21)
Enter,
22)
Select, Cal, enter
23)
Select, Date, enter
24)
Use the “Back Space” key to delete all the characters in the “Date” field and
then press the Enter key
25)
Press Back Key
26)
Select, Save, enter
27)
Select, Save, enter
28)
Note: The unit will display: “Cal changes have been saved.
29)
Press Menu Key
30)
Press Meter Key
31)
Repeat for all modules in the mainframe
Calibration Date is now cleared
Memory sanitization
This memory contains model number, serial number, calibration data and calibration date,
therefore sanitization is not applicable, and there is no provision for sanitizing this memory.
Memory removal
EEPROM is located on module main PCA
Memory validation
N/A