92
Keysight N5990A-155 DisplayPort User Guide
5
Sink Calibration Procedures
Data Rate Deviation Test
Purpose and Method
This test defines the data rate limits where the device can still work.
This procedure adjusts the data rate until the target BER specified in the
CTS cannot be sustained by the DUT. First, it increases the original data
rate value by a deviation step size (by default, 1 ppm); until it reaches the
maximum deviation value. Afterward, it repeats the same procedure but
decreases the original data rate by the deviation step size.
Connection Diagram
Figure 39
Connection Diagram for Receiver Tests
Parameters in Expert Mode
The parameters used in expert mode for this calibration are:
• Bit Error Rate Testing
•
Link Quality Pattern Hint
•
Allowed Number of Errors