Installation and Service Guide N5251-90001
6-15
PNA Series Microwave Network Analyzer System
N5251A
Performance Tests and Checks System
Performance Verification
Figure 6-9
Phase for Mismatched Thru
Interpreting the Verification Results
The purpose of the N5251A system verification process is to demonstrate that the system is
making reasonable measurements. There are no hard specifications on the N5251A system,
so there are no pass/fail limits from the system verification measurements. The results allow
the user to compare the actual measured data to the factory measured data for each
verification device.
The example traces on the previous pages show typical differences between measured and
factory data.
Ignore phase differences for S
11
, S
22
, S
33
, and S
44
when comparing measured data to factory
data.
Содержание N5227A
Страница 4: ...iii Installation and Service Guide N5251 90001...
Страница 7: ...Installation and Service Guide N5251 90001 1 1 1 Safety and Regulatory Information...
Страница 15: ...Installation and Service Guide N5251 90001 2 1 2 System Description...
Страница 25: ...Installation and Service Guide N5251 90001 3 1 3 System Installation...
Страница 41: ...Installation and Service Guide N5251 90001 4 1 4 System Specifications...
Страница 49: ...Installation and Service Guide N5251 90001 5 1 5 Maintenance and Support...
Страница 53: ...Installation and Service Guide N5251 90001 6 1 6 Performance Tests and Checks...
Страница 69: ...Installation and Service Guide N5251 90001 7 1 7 Replaceable Parts...