Installation Note N5225-90103
11
Step 3. Remove Some Bottom-Side (Test Set) Cables
CAUTION
Be careful not to damage the center pins of the semirigid cables. Some flexing of the cables
may be necessary but do not over-bend them.
1. Place the analyzer bottom-side up on a flat surface.
2. Remove the following cables. To see an image showing the location of these cables, click the Chapter 6
bookmark “Bottom RF Cables, Standard 2-Port Configuration, Option 201” in the PDF Service Guide
1
.
These cables may be discarded - they will not be reinstalled.
• W12 (N5245-20109) A29 port 1 receiver coupler to W11
• W18 (N5245-20111) A32 port 2 receiver coupler to W17
• W19 (N5245-20039) A29 test port 1 receiver coupler to front-panel Port 1 SOURCE OUT
• W21 (N5245-20120) A29 port 1 receiver coupler to A37 reference mixer switch
• W31 (N5245-20040) A32 test port 2 receiver coupler to front-panel Port 2 SOURCE OUT
• W35 (N5245-20121) A32 port 2 receiver coupler to front-panel REF 2 SOURCE OUT
• W37 (N5245-20041) Front-panel Port 1 RCVR A IN to A27 mixer brick (A)
• W40 (N5245-20042) Port 2 RCVR B IN to A27 mixer brick (B)
• W46 (N5245-20011) Front-panel REF 2 RCVR R2 IN to A27 mixer brick (R2)
These cables must be saved - they will be reinstalled.
• W41 (N5245-20006) A37 reference mixer switch to front-panel REF 1 SOURCE OUT
• W42 (N5245-20007) REF 1 RCVR R1 IN to A37 reference mixer switch
• W36 (N5245-20155) Front panel jumpers (quantity = 6)
NOTE
: Remove and save the cable guards for the front panel jumpers.
1. See