Getting Started 

Guide 

Keysight 
N4960A 
Serial BERT  
17 and 32 Gb/s 

Содержание N4960A

Страница 1: ...Getting Started Guide Keysight N4960A Serial BERT 17 and 32 Gb s ...

Страница 2: ...nology Licenses The hardware and or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government prime contract or subcontract Software is delivered and licensed as Commercial computer software as defined in DFAR 252 227 7014 June 1...

Страница 3: ... altitudes of up to 2000 meters Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range Before Applying Power Verify that all safety precautions are taken The power cable inlet of the instrument serves as a device to disconnect from the mains in case of hazard The instrument must be positioned so that the operator can easily access the power...

Страница 4: ... EMC standard The RCM mark indicates that this product meets EMS Product Safety Requirements and may be imported to Australia and New Zealand This mark indicates compliance with the Canadian EMC regulations ISM 1 A This text denotes the instrument is an Industrial Scientific and Medical Group 1 Class A product China RoHS regulations include requirements related to packaging and require compliance ...

Страница 5: ...2 96 EC marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in WEEE Directive Annex I this product is classed as a Monitoring and Control instrumentation product Do not dispose in domestic household waste To return unwanted products contact your local Keysigh...

Страница 6: ...6 Keysight N4960A Serial BERT 17 and 32 Gb s Getting Started Guide ...

Страница 7: ...Contents 1 N4960A Getting Started Guide 9 1 1 Introduction 9 1 2 Support 10 1 3 General Specifications 10 1 4 Safety and Regulatory 12 1 5 Unpacking 13 1 6 Important Notes 13 1 7 Installation 13 1 8 Connector Care 15 1 9 Returning the N4960A to Keysight Technologies 16 ...

Страница 8: ...N4960A Getting Started Guide 8 Keysight N4960A Serial BERT 17 and 32 Gb s Getting Started Guide ...

Страница 9: ...rm a quick functional check of the product and guide you on where to find more information and support for the N4960A serial BERT 17 and 32 Gb s The N4960A serial BERT controller 17 and 32 Gb s is shipped in a protective box Each shipping box contains N4960A CJ0 or N4960A CJ1 serial BERT controller AC power cord CD ROM which includes o N4960A Serial BERT 17 and 32 Gb s user guide o N4960A Serial B...

Страница 10: ...Serial BERT 17 and 32 Gb s please refer to N4960A Serial BERT 17 and 32 Gb s User Guide on the CD or the product webpage at http www Keysight com find N4960A Technical Support information http www Keysight com find techsupport 1 3 General Specifications Before installing the N4960A serial BERT review the specifications in Table 3 Table 3 Specification considerations before installation Parameter S...

Страница 11: ...4951A N4951B N4952A remote head controller cable length 1 0 m 39 7 in Weight N4960 CJ0 3 2 kg 7 0 lbs N4951A N4952A with cable 0 86 kg 30 3 oz N4951B with cable 1 0 kg 35 3 oz EMC Complies with European EMC Directive 2004 108 EC IEC EN 61326 1 CISPR Pub 11 Group 1 class A AS NZS CISPR 11 ICES NMB 001 This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme a la norme NMB 001 d...

Страница 12: ...chnologies may result in personal injury or loss of life For continued protection against fire hazard replace fuses and or circuit breakers only with same type and ratings The use of other fuses circuit breakers or materials is prohibited To prevent electrical shock disconnect instrument from mains before cleaning Use a dry cloth or one slightly dampened with water to clean the external case parts...

Страница 13: ... on a flat surface away from heat sources Do not block the fans or the exhaust vents on the rear and side panels 3 min clearance Use a 8 lbf in 90 N cm torque wrench when attaching connectors 1 7 Installation 1 Plug the AC power cord into the N4960A CJ0 or N4960A CJ1 serial BERT controller rear panel power socket 2 Plug the other end of the AC power cord into a suitable wall socket 100 240 VAC 50 ...

Страница 14: ...r Det Menu label then press the softkey corresponding to the SEL label 11 In the Err Det Menu position the arrow next to the AutoAlign Menu label then press the softkey corresponding to the SEL label 12 Position the arrow next to the Perform AutoAlign label then press the softkey corresponding to the EXEC label A small clock appears in the display When the clock disappears ensure that the Errors D...

Страница 15: ...emale receptacles may degrade the 2 92 mm female receptacle Inspect the connectors for the following Worn or damaged threads Scratches to mating surface Burrs and loose metal particles Ensure that female contacts are straight and aligned Clean the connectors as described in the following procedure Cleaning connectors with alcohol shall only be done with the instruments power cord removed and in a ...

Страница 16: ...N4960A fails system verification and you cannot correct the problem return the N4960A to Keysight Technologies for repair following the steps shown below 1 Record all symptoms 2 Contact Keysight Technologies at http www Keysight com find assist 3 Use the original packing material or similar packing material to ship the instrument to Keysight Technologies ...