Making Measurements
2
To Measure Battery-Powered Devices
N2820/1A User’s Guide
31
To Measure Battery-Powered Devices
When making measurements on a battery-powered (floating)
device, such as a mobile phone,
always
connect the supplied
ground lead between ground on your device and the probe’s
ground connector as shown in
. Simply snap the end of
the ground lead onto the probe’s connector. Without the ground
connection, the common mode voltage is not guaranteed to be
within the common mode range of the amplifiers.
Figure 11
Supplied Ground Lead on Device
NOTE
Failure to connect the ground lead may result in inaccurate waveforms.