10
Characteristics
Characteristics
Table 1
Characteristics (Not Warranted)
Maximum Input Voltage (CAT I)
Temperature Coefficient
DC Voltage: 15 kV
AC Voltage: 10 kV
rms
Peak Voltage: 30 kV
peak
(< 20 ns pulse width)
Less than 200 ppm/
°
C
Compensation Range
Division Ratio
For input capacitance of 6 pF to 20 pF
1000:1
Bandwidth
Designed For Use In
50 MHz (–3 dB)
POLLUTION DEGREE 2
Input Resistance
Operating Temperature
100 M
Ω
, when terminated by 1 M
Ω
0 °
C to +50
°
C
Cable Length:
Storage Temperature:
2 meter
−20 °
C to +70
°
C
Altitude:
Humidity
Up to 2,000 meters (6,561 ft)
Up to 80
%
relative humidity at +40
°
C
Accuracy:
DC Volts: ±2, 5% (below 10 kV)
DC Volts: ±4% (between 10 kV and 15 kV)
AC Volts: ±5% at 1 kHz
AC Volts: ±3 dB between 1 kHz and 50 MHz