Keysight N109228CA OIF CEI 3.1 Скачать руководство пользователя страница 77

OIF CEI 3.1 Compliance and Debug Application Getting Started

77

Installing N109228CA on a PC Without N1010A FlexDCA

3

Starting the N109228CA OIF CEI 3.1 Compliance Test Application

After you install the N109228CA OIF CEI Compliance Test Application on a PC without the N1010A 

FlexDCA interface, follow the steps below to start the N109228CA OIF CEI 3.1 Compliance Test 

Application:
1 Ensure that the 86100D or N1000A Oscilloscope is added to the Keysight Connection Expert and 

that the connection is active. Refer to 

“Adding 86100D or N1000A to Keysight Connection 

Expert”

 on page 68.

2 From  the  Windows 

Start

 menu, navigate to 

Start>All Programs>Keysight FlexDCA 

Applications>N109228CA OIF CEI 3_1

and click 

Launch 'N109228CA OIF CEI 3_1'

.

3 The N109228CA OIF CEI 3.1 Compliance Test Application banner appears followed by the 

Connect to FlexDCA

 dialog.

Figure 77

N109228CA OIF CEI 3.1 Application launching banner

4 Enter the SICL address or VISA alias, if known, else select the 

Select from Keysight Connection Expert

 

option to find the SICL Address/VISA alias.

5 Click 

OK

 to connect.

The default view of the N109228CA OIF CEI 3.1 Compliance Test Application appears.

Содержание N109228CA OIF CEI 3.1

Страница 1: ...Keysight N109228CA OIF CEI 3 1 Compliance Test Application Getting Started...

Страница 2: ...under its standard com mercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www key sight com find sweula The license set forth in the EULA rep...

Страница 3: ...stalling the N109228CA OIF CEI 3 1 Compliance Test Application 43 Connecting N1010A FlexDCA to 86100D or N1000A 46 Connecting N1010A FlexDCA to N109X Oscilloscope and N107X 52 Starting the N109228CA O...

Страница 4: ...5 Installing the I O Library Suite 86 Installing the N109228CA OIF CEI 3 1 Compliance Test Application 87 Configuring FlexDCA s SCPI Server Setup 90 Connecting N1010A FlexDCA to N109X Oscilloscope and...

Страница 5: ...r 86100D N1010100A F AEYE or 86100DU 401 license and IO Connection Expert installed on it A PC directly connected to the N109X with a USB cable The PC has N109228CA OIF CEI 3 1 software application N1...

Страница 6: ...86117A 86118A H01 N1045A N1045B N1046A 75 85 100 GHz N1055A or 54754A 86117A 86118A H01 N1045A N1045B N1046A 75 85 100 GHz or N1055A 1 86107A any option 86107A any option 1 Clock Data Recovery and Dem...

Страница 7: ...when testing jitter Required for return loss tests Later PNA models with higher frequency range will also work with the N109228CA Required for Very Short Reach VSR tests The instrument must be able t...

Страница 8: ...N109228CA and N1010A FlexDCA on a PC If your licenses allow for installing the software as marked by the red star in the following diagrams go to Chapter 2 Installing N109228CA and FlexDCA on a PC Fig...

Страница 9: ...OIF CEI 3 1 Compliance and Debug Application Getting Started 9 Introduction 1 Figure 2 N109228CA and N1010A FlexDCA installed on PC connected to N1000A DCA X Oscilloscope...

Страница 10: ...1 Introduction 10 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 3 N109228CA and N1010A FlexDCA Installed on PC connected to N1094B DCA M Oscilloscope...

Страница 11: ...stalling the N109228CA on a PC without FlexDCA If your licenses allow for installing the software as marked by the red star in this diagram go to Chapter 3 Installing N109228CA on a PC Without N1010A...

Страница 12: ...1 Introduction 12 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 5 N109228CA Installed on PC connected to N1000A DCA X Oscilloscope...

Страница 13: ...OIF CEI 3 1 Compliance and Debug Application Getting Started 13 Introduction 1 Figure 6 N109228CA Installed on PC Without N1010A FlexDCA connected to N1094B DCA M Oscilloscope via an other PC...

Страница 14: ...Introduction 14 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 7 N109228CA Installed on PC Without N1010A FlexDCA connected to N1094B DCA M Oscilloscope via 86100 DCA X Oscillosco...

Страница 15: ...CEI 3 1 Compliance and Debug Application Getting Started 15 Introduction 1 Figure 8 N109228CA Installed on PC Without N1010A FlexDCA connected to N1094B DCA M Oscilloscope via N1000A DCA X Oscillosco...

Страница 16: ...ation Getting Started Installing N109228CA on the DCA X Oscilloscope If your licenses allow for installing the software as marked by the red star in this diagram go to Chapter 4 Installing N109228CA o...

Страница 17: ...OIF CEI 3 1 Compliance and Debug Application Getting Started 17 Introduction 1 Figure 10 N109228CA Installed on N1000A...

Страница 18: ...1 Introduction 18 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 11 N109228CA Installed on 86100D connected to a DCA M setup...

Страница 19: ...OIF CEI 3 1 Compliance and Debug Application Getting Started 19 Introduction 1 Figure 12 N109228CA Installed on N1000A connected to a DCA M setup...

Страница 20: ...n and the N1010A FlexDCA Software are installed on the same PC and the DCA X or DCA M Oscilloscope is being accessed remotely see Starting the N109228CA OIF CEI 3 1 Compliance Test Application on page...

Страница 21: ...proceed with connecting to the network analyzer The user interface for Keysight Connection Expert 2017 Version 18 0 and higher is slightly different from its previous versions such as Version 17 2 or...

Страница 22: ...Compliance and Debug Application Getting Started Figure 14 Tabs within Keysight Connection Expert 4 Under Add New Instruments Interfaces select LAN instrument Figure 15 Performing Manual Configuration...

Страница 23: ...the Hostname or IP Address field Note that your Hostname entry will be different from what is shown in the image b Under Set Protocol select Socket and enter the port number c Under Verify Connection...

Страница 24: ...Expert b Click the PNA listing on the left pane and copy the SICL address using Ctrl C keys Note that the address on your end will be different from what is shown in the image Figure 17 Copying Networ...

Страница 25: ...as or SICL address field paste the SICL address using Ctrl V keys d Click Get to confirm the connection e Click OK to close the dialog box f The External Instruments field displays the relevant inform...

Страница 26: ...1 Introduction 26 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 20 Launching the Keysight Connection Expert...

Страница 27: ...Connection Expert click the Add button to manually add instruments and interfaces Figure 21 Adding LAN device manually in the Keysight Connection Expert 2017 4 From the drop down list that appears sel...

Страница 28: ...in the Hostname or IP Address field Note that your Hostname entry will be different from what is shown in the image c Under Set Protocol select Socket and enter the port number d Under Verify Connecti...

Страница 29: ...ication Getting Started 29 Introduction 1 7 Copy Network Analyzer SICL Address to the Clipboard a Click the Instruments tab of the Keysight Connection Expert Figure 24 Selecting the connected Network...

Страница 30: ...he SICL address using Ctrl C keys Note that the address on your end will be different from what is shown in the image Figure 25 Copying Network Analyzer s SICL Address 8 Start the N109228CA OIF CEI 3...

Страница 31: ...CA OIF CEI 3 1 application s main window select PNA from the drop down list b Click Connect The Connect to PNA window appears Figure 27 Connecting to the selected Network Analyzer c In the Enter Selec...

Страница 32: ...1 Introduction 32 OIF CEI 3 1 Compliance and Debug Application Getting Started...

Страница 33: ...tion and the FlexDCA application on the PC You must purchase and install a license for the 86100DU 401 Once the license is installed the 86100DU 401 software will be enabled Figure 28 N109228CA and N1...

Страница 34: ...2 Installing N109228CA and N1010A FlexDCA on a PC 34 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 29 N109228CA and N1010A FlexDCA Installed on PC with N1000A DCA X oscilloscope...

Страница 35: ...network analyzer you will not be able to perform the return loss compliance tests NOTE The 86100DU 400 software is required only for testing the CEI 11G SR Short Reach interface when running the Jitt...

Страница 36: ...Oscilloscope 3 Connect the equipment as shown in Figure 28 or Figure 30 and power on all of the instruments Installed Module Configurations Requires N4877A Limited to CEI 6G LR SR and CEI 11G LR MR S...

Страница 37: ...osuite and download the Keysight I O Library Suite 2 Install the I O Library Suite on the PC NOTE Keysight s I O Library Suite is free software that automatically detects instruments connected to your...

Страница 38: ...n an accessible memory device 4 From the Windows Start menu click All Programs Keysight License Manager to launch the Keysight License Manager Figure 31 Launching Keysight Connection Expert from the S...

Страница 39: ...of the Keysight License Manager window Figure 32 Installing license file on Keysight License Manager version 5 1 above 2 Click Install License File The Select License File s to Install on PC Hostname...

Страница 40: ...ve 5 Close the Keysight License Manager To install the N1010A FlexDCA s license using Keysight License Manager version 5 0 or earlier 1 From the Keysight License Manager window s main menu click File...

Страница 41: ...Getting Started 41 Installing N109228CA and N1010A FlexDCA on a PC 2 2 The Install License File s dialog box appears Figure 36 Selecting license file from the folder 3 Navigate to the folder where th...

Страница 42: ...d Debug Application Getting Started 4 Select the license and click Open The license files are installed and an entry is displayed in the Keysight License Manager Figure 37 Licenses displayed in Keysig...

Страница 43: ...86100DU 401 or N109x 401 license Save the license file that you received from Keysight in a folder on the local disk or in an accessible memory device 5 If the FlexDCA is running on your PC close Flex...

Страница 44: ...the Keysight License Manager window Figure 39 Installing license file on Keysight License Manager version 5 1 above 2 Click Install License File The Select License File s to Install on PC Hostname app...

Страница 45: ...l or right click within the left pane of the Keysight License Manager window and select Install Figure 41 Installing license file on Keysight License Manager version 5 0 or earlier 2 The Install Licen...

Страница 46: ...entry Alternatively you can obtain the computer name from the Windows System properties Figure 43 86100D Oscilloscope s hostname 4 Record the Computer Name corresponds to hostname as you will need th...

Страница 47: ...t FlexDCA Keysight FlexDCA to launch the N1010A FlexDCA Software Figure 44 Launching the N1010A FlexDCA Software on PC NOTE If you are accessing the 86100D Oscilloscope s FlexDCA interface remotely on...

Страница 48: ...nfiguration window is displayed automatically Click Connect if the 86100D Oscilloscope s Hostname is already displayed else click Setup Figure 45 Extended Module Configuration dialog box for connectin...

Страница 49: ...On the FlexDCA Connection Setup dialog box a Ensure that LAN is selected in the Connect Using area b Enter the 86100D s Hostname that you recorded in step 4 Alternatively you may also enter the IP Ad...

Страница 50: ...y the FlexDCA Connection Setup dialog box displays status as Connected Figure 47 86100D Oscilloscope connection status on the FlexDCA Connection Setup 10 Click Close to exit the FlexDCA Connection Set...

Страница 51: ...ramming Tools SCPI Server Setup to open the dialog box 13 Confirm that the Sockets and HiSLIP interfaces are selected Enabled as shown in Figure 49 14 Close the dialog box Figure 49 FlexDCA s SCPI Ser...

Страница 52: ...d Module Configuration window is displayed automatically If the Extended Module Configuration window does not appear from the FlexDCA interface s main menu click Setup Configure Extended Modules 3 Con...

Страница 53: ...on page 77 where the procedure to start the application directly that is without launching the N1010A FlexDCA Software is described Access the N109228CA OIF CEI 3 1 Compliance Test Application using...

Страница 54: ...the FlexDCA s main menu click Apps Automated Test Apps N109228CA OIF CEI 31 4 The N109228CA OIF CEI 3 1 Compliance Test Application banner appears Afterwards the default view of the N109228CA OIF CEI...

Страница 55: ...ebug Application Getting Started 55 Installing N109228CA and N1010A FlexDCA on a PC 2 Connecting to the Network Analyzer Refer to Connecting to External Instruments on page 21 for connecting to the ne...

Страница 56: ...86108A B module Refer to Table 2 on page 36 On the 86100D click Setup Configure Extended Modules In the dialog box perform the following two steps 1 Configure Slot 8 to be the N4877A 2 Remove any oth...

Страница 57: ...when you are installing the N109228CA OIF CEI 3 1 application on a PC but you are not installing the N1010A FlexDCA application on the same PC You must purchase and install a license for the 86100DU...

Страница 58: ...ling N109228CA on a PC Without N1010A FlexDCA 58 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 55 N109228CA Installed on PC Without N1010A FlexDCA connected to N1000A DCA X Oscil...

Страница 59: ...liance and Debug Application Getting Started 59 Installing N109228CA on a PC Without N1010A FlexDCA 3 Figure 56 N109228CA Installed on PC Without N1010A FlexDCA connected to N1094B DCA M Oscilloscope...

Страница 60: ...on a PC Without N1010A FlexDCA 60 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 57 N109228CA Installed on PC Without N1010A FlexDCA connected to N1094B DCA M Oscilloscope via 86...

Страница 61: ...ou do not connect a network analyzer you will not be able to perform the return loss compliance tests NOTE The 86100DU 400 software is required only for testing the CEI 11G SR Short Reach interface wh...

Страница 62: ...Oscilloscope 3 Connect the equipment as shown in the connection diagrams above and power on all of the instruments Installed Module Configurations Requires N4877A Limited to CEI 6G LR SR and CEI 11G L...

Страница 63: ...find iosuite and download the Keysight I O Library Suite 2 Install the I O Library Suite on the PC NOTE Keysight s I O Library Suite is free software that automatically detects instruments connected...

Страница 64: ...f not already done purchase an 86100DU 401 or N109X 401 license Save the license file that you received from Keysight in a folder on the local disk or in an accessible memory device 4 If the N1010A Fl...

Страница 65: ...f the Keysight License Manager window Figure 60 Installing license file on Keysight License Manager version 5 1 above 2 Click Install License File The Select License File s to Install on PC Hostname a...

Страница 66: ...all or right click within the left pane of the Keysight License Manager window and select Install Figure 62 Installing license file on Keysight License Manager version 5 0 or earlier 2 The Install Lic...

Страница 67: ...86100D N1000A Oscilloscope or the PC hosting the FlexDCA software launch the Keysight FlexDCA application 2 Click Tools SCPI Programming Tools SCPI Server Setup 3 In the SCPI Server Setup dialog confi...

Страница 68: ...different from its previous versions such as Version 17 2 or earlier Connection using Keysight Connection Expert v17 2 or earlier 1 On the 86100D Oscilloscope launch the Keysight FlexDCA application...

Страница 69: ...Keysight Connection Expert to launch Connection Expert Refer to Keysight IO Libraries Suite for information on how to install the Keysight Connection Expert Figure 66 Launching the Keysight Connectio...

Страница 70: ...in the Hostname or IP Address field Note that your Hostname entry will be different from what is shown in the image b Under Set Protocol select Socket and enter the port number c Under Verify Connecti...

Страница 71: ...s tab of the Keysight Connection Expert 12 Click the 86100D Oscilloscope s listing on the left pane A check mark encircled in green indicates that the Oscilloscope is connected to your PC remotely on...

Страница 72: ...scroll down to view the Computer Name entry Alternatively you can obtain the computer name from the Windows System properties Figure 71 86100D Oscilloscope s hostname 4 Record the Computer Name corre...

Страница 73: ...ght Connection Expert 2017 7 In Keysight Connection Expert 2017 click the Add button to manually add instruments and interfaces Figure 73 Adding 86100D Oscilloscope manually in the Keysight Connection...

Страница 74: ...in step 1 in the Hostname or IP Address field Note that your Hostname entry will be different from what is shown in the image c Under Set Protocol select Socket and enter the port number d Under Veri...

Страница 75: ...s tab of the Keysight Connection Expert 12 Click the 86100D Oscilloscope s listing on the left pane A check mark encircled in green indicates that the Oscilloscope is connected to your PC remotely on...

Страница 76: ...ded Module Configuration window is displayed automatically If the Extended Module Configuration window does not appear from the FlexDCA interface s main menu click Setup Configure Extended Modules 3 C...

Страница 77: ...at the connection is active Refer to Adding 86100D or N1000A to Keysight Connection Expert on page 68 2 From the Windows Start menu navigate to Start All Programs Keysight FlexDCA Applications N109228...

Страница 78: ...ndow of the N109228CA OIF CEI 3 1 Compliance Test Application NOTE If you click the View N109228CA OIF CEI 3_1 in Demo Mode option under Start All Programs Keysight FlexDCA Applications N109228CA OIF...

Страница 79: ...ug Application Getting Started 79 Installing N109228CA on a PC Without N1010A FlexDCA 3 Connecting to the Network Analyzer Refer to Connecting to External Instruments on page 21 for connecting to the...

Страница 80: ...an 86108A B module Refer to Table 4 on page 62 On the 86100D click Setup Configure Extended Modules In the dialog box perform the following two steps 1 Configure Slot 8 to be the N4877A 2 Remove any o...

Страница 81: ...ow the instructions in this chapter when you are installing both the N109228CA application and the FlexDCA application on the 86100D or N1000A oscilloscope You must purchase and install a license for...

Страница 82: ...4 Installing N109228CA on the oscilloscope 82 OIF CEI 3 1 Compliance and Debug Application Getting Started Figure 81 N109228CA Installed on N1000A...

Страница 83: ...OIF CEI 3 1 Compliance and Debug Application Getting Started 83 Installing N109228CA on the oscilloscope 4 Figure 82 N109228CA Installed on 86100D connected to a DCA M setup...

Страница 84: ...you will not be able to perform the return loss compliance tests NOTE The 86100DU 400 software is required only for testing the CEI 11G SR Short Reach interface when running the Jitter Transfer Bandw...

Страница 85: ...oscope 3 Connect the equipment as shown in the relevant diagrams above and turn on all of the instruments Installed Module Configurations Requires N4877A Limited to CEI 6G LR SR and CEI 11G LR MR SR T...

Страница 86: ...t I O Library Suite 2 Install the I O Library Suite on the same 86100D or N1000A where you will install the N109228CA NOTE Keysight s I O Library Suite is free software that automatically detects inst...

Страница 87: ...or in an accessible memory device 3 If not already done purchase an 86100DU 401 or N1010100A license Save the license file that you received from Keysight in a folder on the local disk or in an access...

Страница 88: ...Keysight License Manager window Figure 85 Installing license file on Keysight License Manager version 5 1 above 2 Click Install License File The Select License File s to Install on PC Hostname appear...

Страница 89: ...right click within the left pane of the Keysight License Manager window and select Install Figure 87 Installing license file on Keysight License Manager version 5 0 or earlier 2 The Install License F...

Страница 90: ...s FlexDCA to start FlexDCA 2 In FlexDCA click Tools SCPI Programming Tools SCPI Server Setup to open the dialog box 3 Confirm that the Sockets and HiSLIP interfaces are selected Enabled as shown in Fi...

Страница 91: ...odule Configuration window is displayed automatically If the Extended Module Configuration window does not appear from the FlexDCA interface s main menu click Setup Configure Extended Modules 3 Config...

Страница 92: ...follow the steps below to start the N109228CA OIF CEI 3 1 Compliance Test Application 1 On the Oscilloscope launch the Keysight FlexDCA application 2 From the FlexDCA s main menu click Apps Automated...

Страница 93: ...d Debug Application Getting Started 93 Installing N109228CA on the oscilloscope 4 Connecting to the Network Analyzer Refer to Connecting to External Instruments on page 21 for connecting to the networ...

Страница 94: ...module Refer to Table 6 on page 85 On the 86100D or N1000A click Setup Configure Extended Modules In the dialog box perform the following two steps 1 Configure Slot 8 to be the N4877A 2 Remove any ot...

Отзывы: