Keysight N109228CA OIF CEI 3.1 Compliance and Debug Application
Getting Started
1
Introduction
Use the instructions in this Guide to install and set up the N109228CA OIF CEI 3.1 Compliance Test
Application. Once the procedures in this document are completed, you can begin compliance testing.
Information on using the application is included in the application’s Online Help.
Your licenses determine where you install the N109228CA and FlexDCA applications. The
N109228CA license allows you to install the application on either a specific PC or a specific 86100D
or N1000A DCA-X instrument. This selection is made at the time you redeem your licenses.
Installing the N109228CA OIF CEI 3.1 Compliance Test Application
1 Select from one of the three different test-equipment configurations shown in the following
pages. The configuration that you select depends on where the N109228CA OIF CEI 3.1
Compliance Test Application is installed, as determined by the licenses, and whether you are
using the N1010A FlexDCA application.
2 After locating the figure that matches your setup, follow the recommended procedure.
Configuration
DCA-X Supported Configurations
DCA-M Supported Configurations
N109228CA on a PC with
FlexDCA
A PC attached to N1000A or 86100D
DCA-X oscilloscope (in Legacy
configuration) over LAN. The PC has
N109228CA OIF CEI 3.1 software
application, N1010A FlexDCA,
86100DU-400 for 86100D, N1010100A
(F-AEYE) or 86100DU-401 license, and IO
Connection Expert installed on it.
A PC directly connected to the N109X with a USB cable. The PC has N109228CA OIF
CEI 3.1 software application, N1010A FlexDCA, N109x-401 or N1010100A License,
and IO Connection Expert installed on it.
N109228CA on a PC without
FlexDCA
A PC attached to N1000A or 86100D
DCA-X oscilloscope (in Standard
configuration) over LAN. The PC has
N109228CA OIF CEI 3.1 software
application, 86100DU-400 for 86100D,
N1010100A (F-AEYE) or 86100DU-401
licenses, and IO Connection Expert
installed on it.
Configuration 1
: Connect the primary test system PC to a low-cost modern PC via
LAN. The FlexDCA interface resides on the second PC. The second PC is then
connected to the N109x via USB. This eliminates most issues of compatibility
between an existing test system PC and the N109X hardware and can greatly simplify
converting an 86100D system to an N109X system.
Configuration 2
: An N1000A or 86100D mainframe connected to the N109X with a
USB cable. The N1000A or 86100D has FlexDCA installed on it. The N1000A or
86100D can then be controlled via GPIB or LAN by a PC that has N109228CA
installed on it.
The PC with N109228CA has N109x-401 or N1010100A License and IO Connection
Expert installed on it.
N109228CA on the
oscilloscope
N1000A or 86100D DCA-X oscilloscope in
Standard configuration. The oscilloscope
has N109228CA OIF CEI 3.1, FlexDCA, and
IO Connection Expert installed on it.
An N1000A or 86100D mainframe connected to the N109X with a USB cable. The
N1000A or 86100D has FlexDCA and N109228CA installed on it.