Keysight M9185A PXI D/A Converter
M9185A Startup Guide
25
4
Read the voltmeter. It should display 16 V with the
accuracy as specified in the specifications.
5
Now configure the module to output 20 mA from the
same channel and change the connection of the HI pin to
HI input of the ammeter.
6
Read the ammeter. It should display 20 mA with the
accuracy as specified in the specifications.
7
Repeat
for each channel for a complete
module verification.
Figure 14
Connections For M9185A verification
If a problem is found
If a problem is found, perform the following checks:
1
Verify that all relevant hardware is turned on.
2
Verify that the M9185A paths are correctly set from the
SFP and that the multimeter is set to the proper voltage
measurement range and that cable is properly connected.
3
If the problem persists, refer to
for details on sending the module to
Keysight Technologies for service.
CHx HI
CHx LO
HI
LO
N O T E
For current measurement, change the HI connection to that of ammeter.