Keysight M8100A Series Arbitrary Waveform Generators Getting Started Guide
21
Introduction
1
Table 7
Insertion/Extraction and Retaining
M8008A Front Panel Connectors
Table 8
M8008A Front Panel Connectors
Component
Description
Retaining screws
The screws on both ends of the module are used to retain the module tightly
inside the M9505A AXIe Chassis slot once you have fully placed it inside the
chassis. To remove the module, you first need to loosen these screws ensuring
that these screws disengage completely.
Module Insertion/Extraction
Handles
The handles on both sides of the module to insert or eject the module from the
slot of the M9505A AXIe Chassis.
CAUTION
The inputs of the M8008A module are sensitive to static electricity.
Therefore, take necessary anti-static precautions, such as wearing a
grounded wrist strap, to minimize the possibility of electrostatic damage.
Connector
Description
Sync In
Reserved for future use
Sync Out A, B, C, D
This output is used to synchronize two or more M8199A modules to a common system
clock. It is connected to the Sync In of the M8199A module.
Sample Clk Out 1 and
Sample Clk Out 2
Sample clock outputs, connected to "Sample Clk In" on up to four M8199A modules.
Each pair of sample clock outputs can be turned on and off independently
Sys Trig In A/B
Reserved for future use
Ref Clk In
This input allows locking the clock frequency to an external 10 or 100 MHz reference
clock
Ref Clk Out
The reference clock output is used to provide a 10 MHz or 100 MHz reference clock to
the DUT or other test equipment.
Ref Clk Out 16G
The reference clock output is used to provide an 8 GHz … 16 GHz reference clock to the
DUT or other test equipment. It can be used as a precision time-based reference signal
for DCA sampling oscilloscopes.