Operation
2
Keysight M1970 Series User’s Guide
27
Operation
The spectrum analyzer performs an LO alignment using a detector in the mixer to set the LO Power. During
the alignment the yellow Busy LED turns on. Once the alignment completes, the mixer is ready to make
calibrated measurements since the conversion loss values stored in the mixer are automatically loaded to the
signal analyzer.
Table 2-1
Equipment setup (continued)for M1970 Series mixers and N9030A, N9020A, and
N9010A signal analyzers (Option EXM) (continued)
Step
Action
Notes
EXT MIXER
SMA cable
LO/IF
USB cable
Signal input to
waveguide
connector
Table 2-2
Operation of M1970 Series mixers and N9030A, N9020A, and N9010A signal
analyzers (Option EXM)
Step
Action
Notes
Tune the
analyzer
a
Apply a signal to the mixer
input.
—
b
Press
FREQ Channel
and
enter a center frequency or a
start and stop frequency.
Multiple responses may appear on screen. Most of these responses are
images or multiples of the mixing process and not true signals.
Содержание M1970 Series
Страница 8: ...8 Keysight M1970 Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 12: ...12 Keysight M1970 Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 14: ...14 Keysight M1970 Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 22: ...1 Overview 22 Keysight M1970 Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 31: ...Operation 2 Keysight M1970 Series User s Guide 31 Figure 2 1 External mixer setup screen...
Страница 38: ...2 Operation 38 Keysight M1970 Series User s Guide Figure 2 2 External mixer setup screen...
Страница 40: ...2 Operation 40 Keysight M1970 Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Страница 52: ...4 M1970 Series Specifications 52 Keysight M1970 Series User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...