From the front panel:
To unsecure the instrument:
1. Press
Utilities
>
Instr. Setup
>
Calibration
.
2. Use
Keyboard
to enter the security passcode and press
Apply
.
3. Press
State
Login
To change passcode:
1. Press
Utilities
>
Instr. Setup
>
Calibration
>
Change Passcode
.
2. Use
Keyboard
to enter the current passcode and your desired passcode.
3. Press
Change
.
From the remote interface:
To unsecure the instrument:
CAL:SEC:STAT 0, EDU36311A
To change a new passcode to K_EDU36311A:
CAL:SEC:CODE K_EDU36311A
To change a forgotten passcode to a new passcode, you can turn on the instrument after shorting CAL
SECURE jumper J1 on the front panel board as shown below and send CAL:SEC:CODE <code> to change the
passcode.
Calibration count
The instrument counts the number of times it has saved calibration data. Your instrument was calibrated at the
factory; when you receive your instrument, read and record the initial count. You can only read the calibration count
by sending the CAL:COUNt? query, and the calibration count is not change by a power cycle or *RST.
If Auto Save is enabled, the count increments whenever a calibration point is calibrated successfully. To avoid double
counting, do not manually save the count with Auto Save enabled.