6 List Sequence Measurements
List Sequence Measurements
Notes
backwards compatibility for the Swept SA measurement. For all other commands/queries which
honor FORMat:DATA, if INT,32 is sent the analyzer will behave as though it were set to REAL,32.
Initial S/W Revision
Prior to A.02.00
The specs for each output type follow:
ASCii - Amplitude values are in ASCII, in the current Y Axis Unit, one ASCII character per digit, values
separated by commas, each value in the form:
SX.YYYYYEsZZ
Where:
S = sign (+ or -)
X = one digit to left of decimal point
Y = 5 digits to right of decimal point
E = E, exponent header
s = sign of exponent (+ or -)
ZZ = two digit exponent
REAL,32 - Binary 32-bit real values in the current Y Axis Unit, in a definite length block.
REAL,64 - Binary 64-bit real values in the current Y Axis Unit, in a definite length block.
Format Data: Byte Order (Remote Command Only)
This command selects the binary data byte order for data transfer and other queries. It controls whether
binary data is transferred in normal or swapped mode. This command affects only the byte order for setting
and querying trace data for the :TRACe[:DATA], TRACe[:DATA]? , :CALCulate:DATA[n]? and
FETCh:SANalyzer[n]? commands and queries.
By definition any command that says it uses FORMat:DATA uses any format supported by FORMat:DATA.
The NORMal order is a byte sequence that begins with the most significant byte (MSB) first, and ends with
the least significant byte (LSB) last in the sequence: 1|2|3|4. SWAPped order is when the byte sequence
begins with the LSB first, and ends with the MSB last in the sequence: 4|3|2|1.
Remote Command
:FORMat:BORDer NORMal|SWAPped
:FORMat:BORDer?
Preset
NORMal
Initial S/W Revision
Prior to A.02.00
294
Sequence Analyzer User's & Programmer's Reference
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...