3 Input/Output Functions
Input/Output
Restore Input/Output Defaults
This selection causes the group of settings and data associated with the Input/Output key to be a reset to
their default values. In addition, when a Source is installed, licensed and selected, Restore Input/Output
defaults will initiate a Source Preset.
This level of Restore System Defaults does not affect any other system settings or mode settings and does
not cause a mode switch. All the features described in this section are reset using this key, including Input
Corrections and Data (described in the Corrections section).
Key Path
Input/Output
Example
:SYST:DEF INP presets all the Input/Output variables to their factory default values.
Notes
Refer to the Utility Functions for information about Restore System Defaults and the complete
description of the :SYSTem:DEFault INPut: command.
Initial S/W Revision
Prior to A.02.00
Corrections
This key accesses the Amplitude Corrections menu.
Amplitude Corrections arrays can be entered, sent over SCPI, or loaded from a file. They allow you to
correct the response of the analyzer for various use cases. The X-series supports four separate
Corrections arrays, each of which can contain up to 2000 points. They can be turned on and off individually
and any or all can be on at the same time.
Trace data is in absolute units and corrections data is in relative units, but we want to be able to display
trace data at the same time as corrections data. Therefore we establish a reference line to be used while
building or editing a Corrections table. The reference line is halfway up the display and represents 0 dB of
correction. It is labeled “0 dB CORREC”. It is drawn in blue.
Corrections data is always in dB. Whatever dB value appears in the correction table represents the
correction applied to that trace at that frequency. So if a table entry shows 30 dB that means we ADD 30
dB to each trace to correct it before displaying it.
In zero span, where the frequency is always the center frequency of the analyzer, we apply the
(interpolated) correction for the center frequency to all points in the trace. In the event where there are
two correction amplitudes at the center frequency, we apply the first one in the table.
Note that the corrections are applied as the data is taken; therefore, a trace in View (Update Off) will not be
affected by changes made to the corrections table after the trace is put in View.
On the RF Input/Output panel, there are two full-duplex RF ports (RFIO1 and RFIO2), RF Input and RF
Output. When RF Input is selected, it will correspond to one input port from two half-duplex RF ports(RFIO3
and RFIO4), and when RF Output is selected, it will correspond to one output port from two half-duplex RF
ports(RFIO3 and RFIO4). So there are 8 sets of corrections in all that can be applied to the RF ports. Ports
cannot share the same set of corrections but a single port can have multiple corrections applied to it. The
correction data is applied to incomming signals as well as transmitted signals and is in the form of a list of
spot frequencies and amplitude correction levels.
Sequence Analyzer User's & Programmer's Reference
101
Содержание E6650A EXF
Страница 34: ...1 About the Test Set Mouse and Keyboard Control 34 Sequence Analyzer User s Programmer s Reference...
Страница 94: ...2 Programming the Test Set Common Commands 94 Sequence Analyzer User s Programmer s Reference...
Страница 129: ...4 Mode Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 129...
Страница 145: ...5 System Functions Undefined variable Primary ProductName Sequence Analyzer User s Programmer s Reference 145...
Страница 206: ...5 System Functions System 206 Sequence Analyzer User s Programmer s Reference...
Страница 214: ...6 List Sequence Measurements File File See File on page 146 214 Sequence Analyzer User s Programmer s Reference...
Страница 474: ...6 List Sequence Measurements Mode Mode See Mode on page 130 474 Sequence Analyzer User s Programmer s Reference...
Страница 478: ...6 List Sequence Measurements Print Print See Print on page 151 478 Sequence Analyzer User s Programmer s Reference...
Страница 642: ...6 List Sequence Measurements System System See System on page 152 642 Sequence Analyzer User s Programmer s Reference...