36
Impedance Analyzer
2
Post Repair Procedure
Post Repair Procedures
lists the required procedures that must be performed after
the replacement of an assembly. These are the recommended
minimum procedures to ensure that the replacement is
successfully completed.
License file
The license code for the frequency option () and generation ID
(G01) must be installed in E:/license, or the error message will be
displayed.
The license code can be downloaded from the Keysight Software
Manager (KSM) or Keysight Software Licensing (KSL).
Table 5
Post Repair Procedures
Replaced
Assembly or Part
Required Adjustment Correction
Constants (CC)
Verification
A11 Source
Board
Perform the required adjust-
ments using “A21 Source Board”
in Spot Adjustment of the pro-
gram.
Troubleshooting Using Diagnos-
Perform all of the Performance
Test.
A6 Receiver
Board
Perform the required adjust-
ments using “A6 Receiver Board”
in Spot Adjustment of the pro-
gram.
Troubleshooting Using Diagnos-
Perform all of the Performance
Test.
A9 DC Bias
Board
Perform the required adjust-
ments using “A9 DC Bias Board”
in Spot Adjustment of the pro-
gram.
Troubleshooting Using Diagnos-
Perform all of the Performance
Test.
Test Head
Perform the required adjust-
ments using “Test Head” in Spot
Adjustment of the program.
Troubleshooting Using Diagnos-
Perform all of the Performance
Test.
A51 DSP Board
Perform the required adjust-
ments using “A51 DSP Board” in
Spot Adjustment of the program.
Troubleshooting Using Diagnos-
Inspect the Booting Process
SSD (Solid State
Drive)
Perform the required adjust-
ments using “HDD” in Spot
Adjustment in the program. Copy
from license file (.lic) under
E:/license
Inspect the Booting Process