Calibration Procedures
3
Keysight E3631A Service Guide
91
230 ! it is off. If not successful, print message to screen and end.
240 !
250 OUTPUT @Psup;"CAL:SEC:STAT OFF, ";Sec_code$
260 OUTPUT @Psup;"CAL:SEC:STAT?"
270 ENTER @Psup;A
280 IF A=1 THEN
290 PRINT TABXY(5,5),"****** Unable to Unsecure the Power supply ******"
300 GOTO 1430
310 END IF
320 !
330 ! Perform the voltage calibration by stepping through one
340 ! power supply at a time. Alert the operator to hook up the
350 ! connection before calibrating.
360 !
370 FOR I=1 TO 3
380 OUTPUT @Psup;"INST:NSEL ";I !Select the output to be calibrated.
390 ! Alert operator to connect lead
400 PRINT TABXY(10,10),"*************************************************"
410 PRINT TABXY(10,11)," Using the annunciator, note the output selected."
420 PRINT TABXY(10,12)," Connect the selected output to the DMM. Observe"
430 PRINT TABXY(10,13)," Polarity."
440 PRINT TABXY(10,14),"*************************************************"
450 PRINT TABXY(10,15),"Press ’C’ to continue, or ’E’ to Exist, then"
460 PRINT TABXY(10,16),"press ’Enter’"
470 Ch$="C"
480 INPUT Ch$
490 IF Ch$="E" OR Ch$="e" THEN GOTO 1420
500 OUTPUT @Psup;"OUTP ON" ! Turn on Power supply output.
Содержание E3631A
Страница 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Страница 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 32: ...1 Specifications 30 Keysight E3631A Service Guide Figure 1 1 Dimensions of Keysight E3631A power supply ...
Страница 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Страница 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Страница 98: ...3 Calibration Procedures 96 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Страница 110: ...4 Theory of Operation 106 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 124: ...6 Component Drawings 120 Keysight E3631A Service Guide E3631 60002 Component Locator for the Main Board Top ...
Страница 125: ...Component Drawings 6 Keysight E3631A Service Guide 121 E3631 60003 Component Locator for the Front Panel ...
Страница 126: ...6 Component Drawings 122 Keysight E3631A Service Guide E3631 60004 Component Locator for the Main Board Bottom ...
Страница 127: ...Component Drawings 6 Keysight E3631A Service Guide 123 Component Locator for the Main Board Top serial MY53xx6xxx ...
Страница 128: ...6 Component Drawings 124 Keysight E3631A Service Guide Component Locator for the Front Panel serial MY53xx6xxx ...
Страница 129: ...Component Drawings 6 Keysight E3631A Service Guide 125 Component Locator for the Main Board Bottom serial MY53xx6xxx ...
Страница 130: ...6 Component Drawings 126 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...