228
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
13
Acquisition Control
This setting only has an effect when sweep speeds are 20 ns/div or faster. At
slower sweep speeds, there are enough sample points in a trigger (acquisition, in
other words) to render a waveform on the display.
Realtime sampling specifies that the oscilloscope produce the waveform display
from samples collected during one trigger event (that is, one acquisition).
Use Realtime sampling to capture infrequent triggers, unstable triggers, or
complex changing waveforms, such as eye diagrams.
When Realtime sampling is on (as in the default setup):
• When less than 1000 samples can be collected in the time spanned by the
screen, a sophisticated reconstruction filter is used to fill-in and enhance the
waveform display.
• If you press the
[Stop]
key, and pan and zoom through the waveform using the
Horizontal and Vertical controls, only the last trigger's acquisition will be
displayed.
When Realtime sampling is off (equivalent time mode):
• Instead of the reconstruction filter, a technique known as random repetitive
sampling is used to build-up and draw one waveform using multiple triggers
(acquisitions).
• Equivalent time sampling mode requires a repetitive waveform with a stable
trigger.
See Also
•
“Realtime Sampling and Oscilloscope Bandwidth"
Realtime Sampling and Oscilloscope Bandwidth
To accurately reproduce a sampled waveform, the sample rate should be at least
2.5 times the highest frequency component of the waveform. If not, it is possible
for the reconstructed waveform to be distorted or aliased. Aliasing is most
commonly seen as jitter on fast edges.
The maximum sample rate for the 3000T X-Series oscilloscopes is 5 GSa/s for a
single channel in a channel pair. Channels 1 and 2 constitute a channel pair, and
channels 3 and 4 constitute another channel pair. For example, the sample rate of
a 4-channel oscilloscope is 5 GSa/s when channels 1 and 3, 1 and 4, 2 and 3, or 2
and 4 are on.
Whenever both channels in a channel pair are on, the sample rate for all channels
is halved. For example, when channels 1, 2, and 3 are on, the sample rate for all
channels is 2.5 GSa/s.
Содержание DSOX3012T
Страница 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Страница 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Страница 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Страница 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Страница 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Страница 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Страница 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Страница 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Страница 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Страница 294: ...294 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 17 Digital Voltmeter and Counter...
Страница 300: ...300 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 18 Frequency Response Analysis...
Страница 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Страница 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Страница 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Страница 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...
Страница 416: ...416 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 25 CAN LIN Triggering and Serial Decode...
Страница 426: ...426 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 26 CXPI Triggering and Serial Decode...
Страница 436: ...436 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 27 FlexRay Triggering and Serial Decode...
Страница 480: ...480 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 30 Manchester NRZ Triggering and Serial Decode...
Страница 510: ...510 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 32 SENT Triggering and Serial Decode...
Страница 520: ...520 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 33 UART RS232 Triggering and Serial Decode...