The LAN Reset button
This recessed button has two functions.
• Pressing the button briefly invokes a preset of the instrument and
restores default measurement settings. This is equivalent to the
programming command:
:SYSTem:PRESet
• Pressing and holding the button for 5 seconds will reset the LAN
parameters to the factory default. This includes removing the password.
This is equivalent to the following programming command:
:SYSTem:COMMunicate:ETHernet:RESet