Setting the Measurement Time Out
The measurement time-out is the time allowed for
each
measurement to
complete. If a measurement
does not
complete before the time out expires,
9.91E37
(not a number) is returned and the display indication is: - - - - - - -. The
sequence continues with the next reading in the sample count.
Specifying a time out prevents the instrument from pausing indefinitely if for some
reason a measurement is unable to complete.
The measurement time out is set with the command:
SYSTem:TIMeout {<time>|MINimum|MAXimum|INFinity|DEFault}
SYSTem:TIMeout? [{MINimum|MAXimum|DEFault}]
(query form)
–
time
is the specified time out in the range of
10 ms
to
2,000s
with
1 ms
resolution. The time out setting can affect reading throughput when multiple
readings are taken during the trigger cycle. Trigger delays, gate times, and
gate delays (
) should be considered when setting an optimum time
out value.
The time out setting is stored in non-volatile memory and
is not
changed
following a reset (
*RST
) or instrument preset (
SYSTem:PREset
or
Preset
key).
When shipped from the factory the measurement time out is set to 1 second.
Specifying a time of
9.9E+37
or sending the
SYSTem:SECurity:IMMediate
command also disables the time out. When the time out is
disabled
, the
instrument will wait
indefinitely
for the measurement to complete.
Содержание 53210A
Страница 1: ...Keysight 53210A 350 MHz RF Counter User s Guide ...
Страница 12: ...12 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...1 Preparation for Use 24 Keysight 53210A User s Guide Figure 1 1 The 53210A Universal Frequency Counter Timer ...
Страница 48: ...1 Preparation for Use 48 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 122: ...4 53210A Input Signal Conditioning 122 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 222: ...8 Instrument Status 222 Keysight 53210A User s Guide Figure 8 1 The 53210A Status System Instrument Locked Global Error ...
Страница 228: ...8 Instrument Status 228 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 240: ...A 53210A Error Messages 240 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 246: ...246 Keysight 53210A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...