Operating Your Instrument
2
Keysight 53147A/148A/149A Operating and Programming Guide
85
2 Press the up-arrow or down-arrow key to decrease or increase the
measurement rate.
You can press these keys as many times as necessary to locate the setting you
want to use. The available resolution settings are
FAST
,
MED
,
SLOW
, and
HOLD
.
3 Press the Enter key to activate your setting and exit the rate-setting mode.
The setting you chose is now in effect. The measurement rate is adjusted
accordingly; you can observe the effect on measurement speed by monitoring
the flash rate of the
Gate
indicator. The rate annunciator is activated when
either
SLOW
or
MEDIUM
is selected for the rate. When you select
HOLD
, the
HOLD
annunciator is activated.
4 If you set the rate to HOLD, press the Reset/Local key to initiate a
measurement.
The instrument computes the measurement and then displays the result (if
averaging is enabled, the instrument displays the
AVERAGING
message while
it computes the measurement). Press the
Reset/Local
key again each time you
want to take an additional measurement.
Hold