3
Programming
232
Keysight 53147A/148A/149A Operating and Programming Guide
[:SENSe] Subsystem
The
[:SENSe]
subsystem commands are divided into several sections. Each
section, or subtree, deals with controls that directly affect
instrument-specific settings and not those related to the signal-oriented
characteristics.
COMMAND
[:SENSe]:AVERage[STATe] . . . <Boolean>
Turns averaging
ON
and
OFF
. When averaging is
ON
, each new valid
measurement result is the average of the number of measurements specified in
the
AVERage:COUNt
command.
COMMENTS
– An
ABORt
command interrupts the averaging and prevents a valid
measurement result.
– Averaging cannot be turned
ON
when the value of
AVERage:COUNt
is one.
Attempting to turn averaging
ON
when
AVERage:COUNt=1
causes an error.
RELATED
FRONT-PANEL
KEYS
Avg
COMMAND
[:SENSe]:AVERage:COUNt . . . <numeric_value>
Specifies the number of measurements to combine when
AVERage:STATe
is
ON
.
RANGE
The acceptable range for the
<numeric_value>
parameter is 1 to 99.
COMMENT
When averaging is
ON
, some devices may automatically set
:COUNt
values in the
TRIGger
subsystem based on the
AVERage:COUNt
value. This is done to ensure
that the
TRIGger
subsystem provides enough triggers for the average.
RELATED
FRONT-PANEL
KEYS
Avg
COMMAND
[:SENSe]:CORRection:CSET:SELect . . . <name>
Selects a power calibration table by name from eight available profiles.
Valid table names:
8481A
,
8481D
,
8482A
,
8485A
,
8487A
,
CUST1
,
CUST2
,
CUST3
COMMENTS
– *RST: OFF