12
Keysight 42030A Four-Terminal Pair Standard Resistor Set
Calibration and Repair
Calibration
2-
Table 2-2
42030A DC Resistance Calibration Uncertainty
Model Number
Nominal Value
Calibration Uncertainty
42031A
1 m
Ω
± 0.1%
42032A
10 m
Ω
± 0.1%
42033A
100 m
Ω
± 0.1%
42034A
1
Ω
± 0.1%
42035A
10
Ω
± 0.03%
42036A
100
Ω
± 0.02%
42037A
1 k
Ω
± 0.02%
42038A
10 k
Ω
± 0.02%
42039A
100 k
Ω
± 0.02%
Table 2-3
42035A 10
Ω
Resistor AC Calibration Uncertainty
Frequency
Calibration Uncertainty
Series Resistance
Reactance
1 MHz
± 0.12%
± 0.010
Ω
2 MHz
± 0.15%
± 0.010
Ω
3 MHz
± 0.20%
± 0.015
Ω
4 MHz
± 0.25%
± 0.025
Ω
5 MHz
± 0.4%
± 0.03
Ω
10 MHz
± 1.0%
± 0.08
Ω
13 MHz
± 1.5%
± 0.10
Ω
Table 2-4
42036A 100
Ω
Resistor AC Calibration Uncertainty
Frequency
Calibration Uncertainty
Series Resistance
Reactance
1 MHz
± 0.12%
± 0.10
Ω
2 MHz
± 0.12%
± 0.10
Ω
3 MHz
± 0.15%
± 0.15
Ω
4 MHz
± 0.15%
± 0.15
Ω