Adjustment Procedures
3
Keysight 3458A Calibration Manual
37
Required Equipment
You will need the following equipment to perform the adjustments:
– A low-thermal 4-terminal short for the offset adjustments (this is typically a
bent piece of copper wire as shown in
– 10 VDC Voltage Standard--Fluke 732A or equivalent (for the DC Gain
Adjustment).
– 10 k
W
Resistance Standard--Fluke 742-10 K or equivalent (for the Resistance
and DC Current Adjustment).
– AC Source-Keysight 3325A Synthesizer/Function Generator or equivalent (for
the AC adjustment).
The resultant accuracy of the multimeter depends on the accuracy of the
equipment used, the thermal characteristics of the short. and the type of cabling
used. We recommend high impedance, low dielectric absorption cables for all
connections.
Содержание 3458A
Страница 1: ...Keysight 3458A Multimeter Calibration Manual ...
Страница 10: ...10 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 14: ...14 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 16: ...16 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 39: ...Adjustment Procedures 3 Keysight 3458A Calibration Manual 39 Figure 3 1 4 terminal short ...
Страница 52: ...3 Adjustment Procedures 52 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 59: ...Performance Verification Tests 4 Keysight 3458A Calibration Manual 59 Figure 4 1 General test procedure ...
Страница 109: ...Command Summary 5 Keysight 3458A Calibration Manual 109 Related commands ACAL CAL SECURE ...
Страница 116: ...A Specifications 116 Keysight 3458A Calibration Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...