1
Keysight 33210A Option 001 External Timebase Reference
8
Keysight 33210A Retrofit Instructions
6
Remove the two screws securing the power supply deck
(top plate) to the chassis.
7
Raise the front of the deck, pull forward, and then lift the deck (with power
supply assembly attached) from the chassis.
CAUTION
Electrostatic discharge (ESD) can damage components on unprotected
circuit boards. Use
static safe
procedures when handling circuit boards.
Avoid high-static areas such as carpeted floors. If an anti-static wrist strap is
available, its use is recommended. Otherwise, touch the metal instrument
chassis to discharge static electricity before removing the timebase
assembly from its anti-static bag. Handle the assembly by its edges, and
avoid touching components.
(Torx T-20)