Keysight 16442B User’s Guide, Edition 3
17
User's Guide
Maintenance
Maintenance
This section provides the following maintenance information:
•
•
“Confirming Interlock Circuit”
•
Cleaning
To maintain high performance, the 16442B test fixture (B1500A-A5F) must be kept
clean. Oil, perspiration, hair, dust, and dirt degrade board insulation, which
increases leakage current and decreases measurement accuracy.
Keysight Technologies recommend the following cleaning procedure:
•
Chassis, lid, inner plate, and socket modules:
1. Make sure that voltage or current is
not
forced to the 16442B.
2. Disconnect all cables from the 16442B.
3. Using lint-free paper, gently wipe the chassis, lid, inner plate, and socket
module. For any area that will not come clean, dip the lint-free paper into
alcohol and wipe the area gently.
•
Connection cables:
1. Make sure that voltage or current is
not
forced to the 16442B.
2. Disconnect the connection cables from the 16442B.
3. Dip lint-free paper into alcohol and gently wipe the metal contacts of the
connection cables.